While the unit is running, press and hold the DOWN button first, then press and hold down the POWER
button. Both buttons must be held down together for five seconds. If the DOWN button is accidentally
released and the POWER button is still being pressed, the unit will turn off. After both buttons have
been held down for five seconds, the display will show "rES". Some of the segments on the E and S will
not be visible until the buttons are released. As soon as the display shows the "rES", both buttons can
be released. The "rES' will be displayed for three seconds and your internal timer will be reset.

MAINTENANCE AND STORAGE

Exterior Pump Case Cleaning Instructions:
1. Clean the exterior case and tubing with a damp (not wet) cloth using mild soap and water solution
once per month or as needed.
WARNING!
Only an authorized technician may open the pump
Before cleaning, unplug power cord from electrical outlet
GARMENT CLEANING/DISINFECTING INSTRUCTIONS:
Disconnect garment from device.
2. Open garment to expose all sides either by separating Velcro type hook and loop or by unzipping
(depending on type of garment).
WARNING!
Do not allow liquids to enter the pump, as this can present an electrical hazard
Always allow the pump to dry before using
Do not use bleach on the pump
3. Cleaning solution should consist of 1/3 cup of laundry detergent per 1 gallon of warm tap water.
Use either a large sink or plastic tub able to hold enough solution (depending on size and quantity
of garments) to completely submerge the garment leaving the latch connector bars out of the wa-
ter.
4. Garment should be soaked for 30 minutes with mild agitation every 5 to 10 minutes while keeping
it below water surface.
5. Thoroughly rinse garment with warm tap water and allow to air dry.
WARNING! Never allow the Latch Connectors to be submerged into the water. If water
enters the inside of the garment, damage may occur to the device.
6. Harder to remove soil on surface of garment may require additional washing by hand with a clean
towel while submerged. Avoid using any abrasive materials such as scrubbing pads or chemicals
that could cause damage to the exterior surface of garment.
7. Re-Submerge garment for 30 minutes (with exception of tubing connectors) in solution consisting
of 1 cup of bleach per 1 gallon of warm tap water, again agitating garment every 5 to 10 minutes
while keeping garment below water surface. Rinse garment thoroughly with warm tap water and
allow to air dry. This completes the disinfecting step.
WARNING! DO NOT place garment in washing machine.
WARNING! DO NOT use the tubing or valves as "handles" for carrying, handing or storing
garment.
9