Cable Data - Siemens SITRANS FM MAG 5000 Operating Instructions Manual

Electromagnetic flowmeters ip67
Hide thumbs Also See for SITRANS FM MAG 5000:
Table of Contents

Advertisement

Totalizer lap
Batch on relay output
Unidirectional mode (forward flow only)
Hold
10.4

Cable data

Description
Electrode or coil cable (standard)
Electrode cable, double shielded (for empty pipe detection or low
conductivity fluids)
Cable kit with standard coil cable and electrode cable double
shielded (also available as low noise cable for MAG 1100 sensor)
Technical data
Basic data
SITRANS FM MAG 5000/6000 IP67
Operating Instructions, 12/2019, A5E02338368-AD
No. of conductors
Sqr. area
Screen
Color code
Outside color
Ext. diameter
Conductor
Isolation material
Batch
Standard cable
(electrode/coil)
3
1.5 mm
2
Yes
Brown, blue, black
Grey
7.8 mm
Flexible CU
PVC
Technical data
10.4 Cable data
Double-shielded cable
(electrode)
3
0.25 mm
2
Double
Brown, blue, black
Grey
8.1 mm
Flexible CU
PVC
75