2.6.4

External Sensor Interface

ONLY an Easidew I.S. Transmitter should be connected if the
HygroPort I.S. is to be used in a hazardous area. The Easidew
!
I.S. must be connected using the I.S. barriers required by CE-
ATEX, FM, or CSA certification. A Kahn Remote Sensor
Interface module is also required.
Highlight EXTERNAL from the Main Menu
by using a combination of the scroll up and
scroll down keys and press the Enter key. The 
External Set-Up Menu, is then displayed, see
Figure 2.11.
Initially
the EXTERNAL TYPE
highlighted.
This field is used to select a
Kahn Easidew I.S. Transmitter.
To enter the parameter value field press the
Enter key. The current setting i.e. NONE, is
then highlighted and can be changed. Use
either the scroll up or scroll down keys to
select EASIDEW and then press the Enter key.
Figure 2.11 illustrates the selection of an
Easidew I.S. transmitter. NOTE: The ZERO
and SPAN limits are automatically added
for this transmitter.
Use the scroll up and scroll down keys to
select each parameter that needs to be
entered in turn and press the Enter key.
A user pressure can also be selected. This
pressure is used in the calculation of dew
point at pressure, moisture content and
absolute humidity. If a user pressure is not
selected then HygroPort I.S. assumes atmospheric pressure.
Table 2.6 details the field values. To quit, press the scroll left key.
Parameter
Parameter variable options
EXTERNAL
Defines use of Easidew I.S. transmitter. ZERO and SPAN (fullscale)
default limits are pre-programmed but can be redefined
TYPE
EXTERNAL
Defines external zero (lower limit). The ZERO default limit is pre-
programmed but can be redefined
ZERO
EXTERNAL
Defines external span (upper limit). The SPAN (full scale) default
limit is pre-programmed but can be redefined
SPAN
USER
Range 0-5000 psig (or equivalent in other units). Used to define
system pressure, for use in calculating moisture parameters
PRESSURE
Issue 6, December 2019
field
Table 2.6
External transmitter selection
17
EXTERNAL TYPE
EXTERNAL ZERO
EXTERNAL SPAN
USER PRESSURE
is
EXTERNAL TYPE
EXTERNAL ZERO
EXTERNAL SPAN
USER PRESSURE
EXTERNAL TYPE
EXTERNAL ZERO
EXTERNAL SPAN
USER PRESSURE
EXTERNAL TYPE
EXTERNAL ZERO
EXTERNAL SPAN
USER PRESSURE
Figure 2.11
Kahn Instruments
NONE
EXTERNAL
N/A
N/A
ATM
Psig
NONE
EXTERNAL
N/A
N/A
ATM
Psig
EASIDEW
EXTERNAL
-100
°C
20
°C
ATM
Psig
EASIDEW
EXTERNAL
-100
°C
20
°C
ATM
Psig
External transmitter