Figure C.1 Current Datalog File Display; Figure C.2 Hygroport I.s. Status Register; Figure C.3 Hygroport I.s. Status Register (Hex 28) - Kahn HygroPort I.S. User Manual

Hygrometer
Table of Contents

Advertisement

HygroPort I.S. User's Manual
Appendix C
Datalog Status Display
Introduction
When the status of the current log file is displayed, see Figure C.1
the current log file display represents the instrument's status at the time that each log
point is recorded.
#
106
26.8
107
26.2
108
25.9
109
25.8
110
25.5
This column shows a hexadecimal representation of the instrument's 16-bit status
register shown below in Figure C.2. Bit 1 is the least significant bit. The four most
significant bits, Bits 13 to 16 are for Kahn Instruments' service use and so the first
character in any HygroPort I.S. ST display will always read 0.
16
15
14
13
N/A
N/A
N/A
N/A
For the example given for each of the log plots shown on in Figure C.1 (Hex 0028), the
output pattern of the status register will be as shown in Figure C.3
16
15
14
13
N/A
N/A
N/A
N/A
Figure C.3
Each of the error bits set high (1) represents a specific status or error condition. In
this example, Bits 4 and 6 are set high indicating that two (status) conditions exist.
Issue 6, December 2019
EXT
ST
INT
-100.0
0028
-90.0
0028
-80.0
0028
-70.0
0028
-60.0
0028
Figure C.1
Current datalog file display
Status register Bit position
12
11
10
9
0/1
0/1
0/1
0/1
Figure C.2
HygroPort I.S. status register
Status register Bit position
12
11
10
9
0
0
0
0
HygroPort I.S. status register (Hex 28)
55
File
CR270309
INT
RH
EXT.
EXT_DP
INTVAL
05 Sec
D#
090327
T#
123606
8
7
6
5
0/1
0/1
0/1
0/1
0/1
8
7
6
5
0
0
1
0
Kahn Instruments
ST
,
the
column in
4
3
2
1
0/1
0/1
N/A
.
4
3
2
1
1
0
0
N/A

Advertisement

Table of Contents
loading

Table of Contents