Kahn S4000 Integrale Installation, Operation And Maintenance Manual

4000 integrale series precision dewpoint hygrometer

Advertisement

Quick Links

SERIES 4000 PRECISION
DEWPOINT HYGROMETER
Integrale
INSTALLATION, OPERATION AND
MAINTENANCE MANUAL
Issue July 2001
KAHN INSTRUMENTS, INC.
885 Wells Road, Wethersfield, CT 06109
Tel: 860-529-8643; Fax: 860-529-1895
E-mail: info@kahn.com
www.kahn.com

Advertisement

Table of Contents
loading

Summary of Contents for Kahn S4000 Integrale

  • Page 1 SERIES 4000 PRECISION DEWPOINT HYGROMETER Integrale INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue July 2001 KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 2: Table Of Contents

    Operating the Automatic Balance Compensation System 4.4.3 Periodic ABC Operation 4.4.4 On Request ABC Operation User Inputs / Outputs 4.5.1 Analog Inputs / Outputs KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 3 4.5.2 Digital Interface Microscope TECHNICAL SPECIFICATIONS General Sensor Control Electronics MAINTENANCE Cleaning the Optical System 6.1.1 Cleaning Procedure OPTIONS RS485 Output Remote Pressure Transducer KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 4: Introduction

    Page 4 INTRODUCTION General Kahn Instruments manufactures a wide range of on-line and portable instruments for the determination of moisture content in air and gases. In addition, the company manufactures humidity calibration systems for industrial and scientific users. Integrale The Series 4000...
  • Page 5: Operating Principle

    Quality System. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 6: Instrument Description

    The black blanking plug may be removed from the sensor and a microscope attached to the sensor assembly to allow the formation of ice crystals to be observed by the operator. Refer to Section 4.6. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 7: Front Panel Controls

    Manually starts a Balance cycle. Refer to Section 2.8. f) ABC STATUS LED Indicates current status of Balance cycle. Refer to Section 2.7 and 2.8. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 8: Rear Panel Controls / Connections

    Overload protection fuse for sensor’s Peltier heat pump. Fuse rating T 3.15 A 250V. i) AC Power Input I.E.C. Socket for AC power supply. Fused at T3.15A 250V. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 9: Operation / Standby Modes

    1. Only one switch may be in the ON position at any time. 2. One switch must be in ON position, otherwise, if an ABC is initiated it will continue indefinitely. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 10: Automatic Balance Compensation On Request

    This function holds the display, analog output and digital output for the defined time period, and is activated whenever a Balance cycle is initiated, i.e. whether manually or automatically. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 11: Clean Optics Warning

    This temperature will be indicated on the Monitor Unit digital display. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 12: Display Operation

    °C and °F. The two digit annunciator will display “F” indicating degrees Fahrenheit. The “UNIT SELECT” switch will scroll through the functions as follows: KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 13: Lower Display

    • Pressure (kPa) • Sample Gas Temperature (°C) • Sample Gas Temperature (°F) The “UNIT SELECT” switch will scroll through the above functions as follows : KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 14 3. If the upper display is indicating %RH, and the lower “UNIT SELECT” push button is pressed, then PPM will be displayed, and the upper display reverts to dewpoint in either °C or °F. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 15: Pressure Measurement

    The pressure is displayed in kPa, and is available as a digital output on the RS232 interface. Refer to Section 4.5.2. The standard pressure range is 0 to 50 psi absolute, but special pressure ranges may be supplied to order. Contact Kahn Instruments Technical Sales Department for further information. Remote Pressure Connection...
  • Page 16: Installation

    Also, it is important that a sufficient flow of ambient air reach the rear of the unit. The ventilation holes on the top of the instrument must not be obstructed. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 17: Connecting The Sample Gas

    P.R.T. probe should be placed in the gas flow, and the probe connected to the connector on the rear of the unit. Further electrical connections for analog or digital I/O may be required, refer to Section 4.5. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 18: Operation

    For example, in a glove box application connect the sensor at the exit of the glove box not at the gas entry point. Minimize dead space in sample lines. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 19: Which Gases To Measure

    If particulates such as desiccant dust or rust are possible, use a particulate in-line filter. Kahn instruments' technical sales staff will be pleased to give advice.
  • Page 20: Operating The Hygrometer

    The instrument will now perform a balance cycle to compensate the ABC system for the current level of contamination of the optical system. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 21: Periodic Abc Operation

    For laboratory use or calibration work it may be desirable to operate the Series 4000 in this mode, to prevent the automatic initiation of the balance cycle causing interference with critical measurements. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 22: User Inputs / Outputs

    Input The “USER I/O” connector located on the Monitor rear panel is a 12 way screw-terminal type and accepts wire pin or bare wire connections. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 23: Digital Interface

    Relative Humidity (%RH) Temperature ( ° C) • Temperature ( ° F) • To initiate data transfer from the S4000 instrument, a request command is required. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 24 72.68 C -42.39 F Additionally, a Balance cycle may be initiated through the RS232 interface. To initiate an ABC the request command is abc<CR> or ABC<CR> KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 25: Microscope

    2 °C. Super-cooled water formation leads to dewpoint values lower than would be seen if ice crystals were formed. If you require further advice on this matter please contact Kahn Instruments' Technical Sales Department.
  • Page 26: Technical Specifications

    9.8 kg (21.6 lb ). Sensor Mirror cooling: Three stage Peltier with forced air cooling. Mirror: Plated copper. -60 to +35 ° Cdp. Dewpoint range: KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 27: Control Electronics

    10 mV / ° C dewpoint Analog: 4-20mA over the range -60 to +30 ° C dewpoint Digital: Data Hold Balance Cycle Optics Alarm KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 28 +/- 0.2 ° C. Gas Temperature Accuracy: PRT Probe Type: Stainless Steel, sheathed, with 2 meter cable. 1/10 DIN class B 3 wire PRT. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 29: Maintenance

    MAINTENANCE Cleaning the Optical System Integrale The Kahn Series 4000 is a precision optical dewpoint hygrometer, which uses a four component optical system to detect condensation of water from a gas sample. This system comprises a light emitting diode (LED), two photo-detectors and a mirror surface.
  • Page 30 However, it is good practice to ensure the cleaning procedure is carried out regularly rather than waiting for the red ABC STATUS LED to illuminate. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...
  • Page 31: Options

    The standard pressure range is 0 to 50 psi absolute, i.e. at / around atmospheric pressure. But particular pressure ranges may be supplied to order. Contact Kahn Instruments’ Technical Sales Department for further information. KAHN INSTRUMENTS, INC.
  • Page 32 Page 32 Integrale APPENDIX 1 S4000 Outline Drawings POWER INCL. FEET KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 33 CURRENT SYSTEM RH 0-10V SENSOR CONTROL OUTPUT RH 4-20mA OUTPUT SAMPLE USER CONTROL DIGITAL INTERFACE TEMP REMOTE TEMPERATURE 10mV/ºC MEASUREMENT OUTPUT TEMP 4-20mA OUTPUT KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com www.kahn.com...
  • Page 34 Dewpointmeter operating in an ambient temperature of +21°C, the maximum depression is –62°C. The upper dewpoint measurement limit is +19°C dewpoint and the lower dewpoint measurement limit is -60°C dewpoint. KAHN INSTRUMENTS, INC. 885 Wells Road, Wethersfield, CT 06109 Tel: 860-529-8643; Fax: 860-529-1895 E-mail: info@kahn.com...