HygroPort I.S. User's Manual
Gas samples
Gases containing entrained solids should be filtered before application to the
instrument. Please consult Kahn Instruments prior to the measurement of any
sample suspected to be corrosive.
Care should be taken with gas mixtures containing potentially condensable
components in addition to water vapor, e.g. oil, to insure that only water vapor
is present in the sample. Once present on the surface of the sensors, oil will
remain, and reduce the sensitivity of the sensor.
apparatus attached to this instrument must be adequate
!
otherwise physical injury to the operator or bystander is
Before disconnecting the HygroPort I.S. from the gas line
!
it is essential to vent the system to atmospheric pressure,
Issue 6, December 2019
POSSIBLE INJURY! The tubing, valves and other
for the maximum pressure which will be applied,
otherwise severe injury could result.
42
possible.
Kahn Instruments