14.3 EMG Interference

1 Whether or not the room is comfortable?
2 Whether or not the patient is nervous?
3 Whether or not the bed space is narrow?

14.4 Baseline Drift

1 Whether or not the electrode placement is instable?
2 Whether or not the electrodes or lead cables are connected reliably?
3 Whether or not the electrodes and patient skin are cleaned and are smeared enough
electric gel?
4 Whether or not it is caused by patient's movement and breath?
5 Whether or not the electrodes or leads are in bad connection?
If you can not clear the interference after having taken all the measures above, use
filter.