VI
1

Specifications

Volume of LOX (typical)
Weight of LOX at 24 psig (166 kPa)
Saturation (typical)
Gaseous Oxygen Equivalent @ 1atm.
and 70°F
Height
Dimensions (Footprint)
Empty Weight
Full Weight
Economizer Pressure
Primary Relief Valve Opening Pres-
sure
Primary Relief Valve Reseat Pressure
Secondary Relief Valve Opening
Pressure
Secondary Relief Valve Reseat Pres-
sure
Normal Evaporation Rate (NER)
(typical)
Selectable Flow Rates
Contents Indicator
Operating Temperature
Storage Temperature
Table 4: Equipment Specifications
HELiOS Plus (H300)
0.38 L
0.9 lb (0.41 kg)
308 L
10.5 in (26.67 cm)
5.78 x 3.68 in (14.68 x 9.34 cm)
2.7 lb (1.22 kg)
3.6 lb (1.63 kg)
1.41-1.59 bar (20.5-23 psi)
≤1.90 bar (≤27.5 psi)
>1.59 bar (>23 psi)
5.24-5.79 bar (76-84 psi)
>4.41 bar (>64 psi)
0.85 lb/day (0.39 kg/day)
Continuous (LPM): .12, .25, .5, .75
Demand: 1, 1.5, 2.0, 2.5, 3, 3.5, 4
Spring Scale
-10°C to 40°C (14°F to 104°F)
95% max. relative humidity
-40°C to 70°C (-40°F to 158°F)
95% max. relative humidity
HELiOS Marathon (H850)
0.84 L
2.0 lb (0.90 kg)
685 L
15.2 in (38.61 cm)
5.78 x 3.68 in (14.68 x 9.34 cm)
3.6 lb (1.63 kg)
5.6 lb (2.54 kg)
1.53-1.71 bar (22.25 - 24.75 psi)
1.79-2.07 bar (26-30 psi)
>1.72 bar (>25psi)
5.24-5.79 bar (76-84 psi)
>4.41 bar (>64 psi)
1.0 lb/day (0.45 kg/day)
Continuous (LPM): 1, 2, 3, 4, 5, 6
Demand: 1.5, 2, 2.5, 3, 4
Spring Scale
-10°C to 40°C (14°F to 104°F)
95% max. relative humidity
-40°C to 70°C (-40°F to 158°F)
95% max. relative humidity
HELiOS Portables Technical Service Manual • PN 20562190 Rev E