ECLIPSE 5 OXYGEN SYSTEM SPECIFICATIONS

Oxygen Concentrator
Dimensions (H x W X D)
Weight
Eclipse 5
Power Cartridge
Flow Settings
Continuous Flow (measured in Liters Per Minute LPM) Pulse
Dose (measured in mL)
Continuous Flow Accuracy
Oxygen Concentration
Maximum System Pressure
Oxygen Output Pressure
Oxygen Concentration
Status Indicator
Nominal Sound Level
3.0 LPM Continuous Flow
3.0 Pulse Dose Mode
Operating Environment
Temperature
Humidity
Storage Environment Temperature
Humidity
Altitude
Nominal Power
0.5 LPM Continuous flow 3 LPM
Continuous Flow
1.0 Pulse Dose Setting
6.0 Pulse Dose Setting Battery
Charging
Nominal Power Cartridge operating time* Continuous
Flow
2.0 LPM
3.0 LPM
Pulse Dose (@ 12 BPM)
Setting 2.0
Setting 6.0
Outlet Pressure
Continuous Flow Indication
Audible Alarm Indicators
19.3 x 12.3 x 7.1 inches
(49.0cm x 31.2cm x 18.0cm)
15.0 pounds
3.4 pounds
0.5 to 3.0 LPM (0.5 liter increments)
Settings 1.0-6.0: 16mL, 24mL, 32mL, 40mL, 48mL, 56mL, 64mL, 72mL, 80mL, 88mL, 96mL
Settings 7-9: 128mL, 160mL, 192mL
+/- 10% or 200ml/min, whichever is greater
87–95.5% for all flow settings
14 psig (96.5 kPa)
5.0 psig (34.5 kPa) nominal
Green Light = Normal Operation
Yellow Light = Warning or Caution, less than 85% ± 3%
Red Light Flashing = Abnormal Operation, less than 70% ± 5% (Note: only for 3-LED models)
48 dB(A)
40 dB(A)
Less than 57 dB while alarming
50º F to 104º F (+10° to 40°C)
10% to 95%, Non-condensing, 82.4°F (28°C) Maximum Dew point
-4º F to 140º F (-20° to 60°C)
Up to 95% Non-condensing
0 – 13,123 feet (0 - 4,000 meters)
48 Watts
145 Watts
45 Watts
95 Watts
50 Watts
2.0 hours
1.3 hours
5.2 hours
3.5 hours
* Note: Times will decrease with higher bolus size, breath rate, ambient temperature, power
cartridge age and use over time.
Nominal: 7.0 psig
Maximum: 14.0 psig
Expressed in liters per minute (LPM)
• Loss of Power/Hot Power Cartridge
• Low Power Cartridge/Warm Power Cartridge
• Low Oxygen Output
• O
Flow Outside Normal Limits
2
• Unit Malfunction
• Low 9V battery (Note: only for 3-LED model)
Oxygen Concentrator Technical Service Manual | PN 20631679 Rev G