  • Page 1 Sprint / Stroller / Hi Flow Stroller User Manual (English)
  • Page 2: Symbols Glossary

    Sprint / Stroller / Hi Flow Stroller Symbols Glossary This device complies with the ISO 7000; Graphical symbols for use on equip- requirements of Directive 93/42/EEC ment—Index and synopsis concerning medical devices. It bears Operating temperature limitation of the CE marking as shown.
  • Page 3: Specifications

    Sprint / Stroller / Hi Flow Stroller Specifications • Mode of Operation: Continuous Flow • Type of Protection Against Electrical Shock: Internally Powered Equipment • Degree of Protection Against Electrical Shock: Type BF Applied Part • Classification According to the Degree of Protection Against Ingress of Water: IP22 – Ordinary Equipment •...
  • Page 4: Warning Information

    Sprint / Stroller / Hi Flow Stroller Warning Information Keep flammable materials away from this equipment. Aerosol sprays, oils and Important: Read this manual thoroughly before grease, including facial creams and pe- operating the Sprint/Stroller/Hi Flow Stroller. troleum jelly, ignite easily and may burn RX Only.
  • Page 5 Sprint / Stroller / Hi Flow Stroller Introduction Controls 1. Electronic Liquid Level Gauge The Sprint, Stroller, and Hi Flow Stroller are in- 2. Electronic Liquid Level Switch tended for the administration of supplemental oxy- 3. Mechanical Liquid Level Gauge gen to the patient in the end user’s home and can...
  • Page 6 Sprint / Stroller / Hi Flow Stroller 3. Position the fill connector in the upright position Quick Disconnect Valve over the stationary unit’s connector. (QDV) Identification Portable Stationary QDV Style Connector Connector CAIRE Side Fill Locking 4. a. If your connector is a push-on:...
  • Page 7 NOTE: For push-on style connector, hold the unit in Portable Release Button (push-on style only) place throughout the fill. 6. Open the vent valve to begin filling your Sprint/ Stroller/Hi Flow Stroller. You should hear some oxygen escaping; do not worry. The hissing sound...
  • Page 8 IN INCREASED FIRE HAZARD. oxygen comfortably. Caution: Should there be any liquid leak- 3. Turn the flow control knob on top of the Sprint/ age from the stationary unit or Sprint/ Stroller/Hi Flow Stroller/Hi Flow Stroller clock- Stroller/Hi Flow Stroller separating the...
  • Page 9 Sprint / Stroller / Hi Flow Stroller 5. Use the following chart as a guideline to deter- mine the length of time the Sprint/Stroller/Hi Flow Stroller will operate: FCV Setting Model Sprint Stroller Hi Flow Stroller Nominal Nominal Sprint Stroller Hi Flow Pos.
  • Page 10 Sprint / Stroller / Hi Flow Stroller • Remove the pad from the cup and wring out NOTE: If the unit is empty, only the red colored band any absorbed moisture. Allow the pad to dry of the scale will be exposed. If this is the case, the completely before replacing it in the cup.
  • Page 11: Troubleshooting

    Sprint / Stroller / Hi Flow Stroller Troubleshooting The following information is intended to help you troubleshoot and solve simple operational problems that you may experience when using your Sprint/Stroller/Hi Flow Stroller. Issue Solution The Portable makes a • Hissing can occur to maintain the correct operating pressure within the Portable.
  • Page 12: Maintenance

    Sprint / Stroller / Hi Flow Stroller Cleaning Standard Maintenance Your service provider is responsible for any mainte- WARNING: CLEAN ONLY AFTER THE UNIT nance that my be required per the technical manual IS EMPTY. of this device. Call your service provider for any •...
  • Page 13 Guidance and Manufacturer’s declaration—electromagnetic emissions The Sprint/Stroller/Hi Flow Stroller is intended for use in the electromagnetic environment specified below. The customer or the user of the Sprint/Stroller/Hi Flow Stroller should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment—guidance...
  • Page 14 Guidance and manufacturers declaration—electromagnetic immunity The Sprint/Stroller/Hi Flow Stroller is intended for use in the electromagnetic environment specified below. The customer or the user of the Sprint/Stroller/Hi Flow Stroller should assure that it is used in such an environment. Immunity test...
  • Page 15 Guidance and Manufacturer’s Declaration—Immunity ME Equipment and ME Systems Guidance and Manufacturer’s Declaration—Immunity The Sprint/Stroller/Hi Flow Stroller is intended for use in the electromagnetic environment specified below. The customer or user of the Sprint/Stroller/Hi Flow Stroller should ensure that it is used in such an environment. Immunity Test...
  • Page 16 Tel/Fax: (55)(11)5535-0989 email: medstar@medstar.com.br Copyright © 2018 CAIRE Inc. CAIRE Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions, specifications and/ |MN236_A\~ or, processes to its products at any time without prior notice and with no further obligation or consequence.