Specifications; Intended Use - CAIRE HELiOS Series User Manual

Hide thumbs Also See for HELiOS Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Specifications

• Mode of Operation: Demand and Continuous
Flow
• Type of Protection Against Electrical Shock:
Internally Powered Equipment
• Degree of Protection Against Electrical
Shock: Type BF Applied Part
• IPX1 Classification According to the Degree
of Protection Against Ingress of Water: Drip
Proof
• Equipment not suitable for use in the
presence of flammable mixtures
HELiOS Plus
(H300)
LOX capacity
0.9 lb (0,41 kg)
Gaseous
308 L
equivalent
capacity
Weight, filled
3.6 lb (1,63 kg)
Weight, empty
2.7 lb (1,22 kg)
Height
10.5" (267 mm) 15.0" (381 mm)
including
handle
Typical use
9 hours
time at 2 LPM
demand mode
Typical use
N/A
time at 2 LPM
continuous
mode
Operating
22 psi (152 kPa) 22 psi (152 kPa)
pressure
Normal
1.0 lb/day
evaporation
(0,45 kg/day)
rate
Standard flow
Continuous:
control range
off, .12, .25, .5,
.75 Demand:
off, 1, 1.5, 2,
2.5, 3, 3.5, 4
Introduction
A liquid oxygen system is designed to provide
supplementary oxygen as prescribed by a
physician. Your liquid oxygen system includes a
HELiOS portable unit and a stationary reservoir.
The reservoir is the larger container that is used
to fill your HELiOS portable. This user manual
contains only the instructions for using the HELiOS
portable. Refer to the user manual supplied with
your reservoir unit for information on its operation.
HELiOS portable liquid oxygen units are designed
to be an ambulatory source of oxygen for an
extended period of time. They are filled with liquid
HELiOS
oxygen from a reservoir unit with a compatible fill
connector. HELiOS portables are available in two
Marathon
different sizes.
The H300 Plus is primarily a conserving device
(H850)
and holds 0.38 Liters of liquid oxygen. It provides
2.0 lb (0,91 kg)
demand (pulse) flow settings 1-4, and limited low
dose continuous flow settings at 0.75 LPM and
685 L
below only.
The H850 Marathon can be used as a conserving
or continuous flow device and holds 0.84 liters of
5.6 lb (2,54 kg)
liquid oxygen. It provides continuous flow from
1-6 LPM and demand flow on settings of 1-4.
3.6 lb (1,63 kg)
22 hours
6.1 hours
0.85 lb/day
(0,39 kg/day)
Continuous: off,
1,2,3,4,5,6
Demand: off,
1.5, 2, 2.5, 3, 4

Intended Use

The CAIRE HELiOS Liquid Oxygen unit is intend-
ed for the administration of supplemental oxygen.
The device is not intended for life support nor does
it provide any patient monitoring capabilities.
HELiOS 300 & 850
H300 and H850 portable units shown.
PN 21033264-C4 C | User Manual
5 - ENG

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Helios plusHelios marathonH300H850

Table of Contents