CAIRE HELiOS Plus Technical & Service Manual page 49

Hide thumbs Also See for HELiOS Plus:
Table of Contents

Advertisement

XI
Troubleshooting & Repair Procedures
Table 8: H850 Flow Tolerance
Flow
Measured Flow
Setting
D1.5
Tail Flow
D2
Tail Flow
D2.5
Tail Flow
D3
Tail Flow
D4
Tail Flow
0
None
C1
Continuous
C2
Continuous
C3
Continuous
C4
Continuous
C5
Continuous
C6
Continuous
NOTE: The values in Tables 7 & 8 are based on ambient
conditions of 70oF (21oC) and 29.4 in Hg (74 mm Hg) and an
operating pressure range of 1.54 to 1.71 bar (22.3 psi to 24.8
psi). Any deviations from these conditions can affect flow
results.
NOTE: If the flow measurements are out of specification,
check the pressure in the unit. A combination of high or low
pressure and the tolerance of the particular flow meter can
result in inaccurate readings.
Specification
Acceptable
(LPM)
Range (LPM)
0.65
0.5 - 0.8
0.75
0.6 - 0.9
1.00
0.85 - 1.15
1.50
1.3 - 1.7
2.00
1.7 - 2.3
0.00
0.00
1.00
0.85 - 1.15
2.00
1.7 - 2.3
3.00
2.6 - 3.4
4.00
3.5 - 4.5
5.00
4.4 - 5.7
6.00
5.3 - 6.8
HELiOS Portables Technical Service Manual • PN 20562190 Rev E