Application Measurements And Calculations - EDAN Acclarix LX4 User Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

Acclarix LX4 Diagnostic Ultrasound System User Manual
3. Press the Set key to place that caliper and add the second half of the caliper pair. Presses
of the Update key toggle which side of the caliper can be moved. Pressing the Set key
completes the measurement and allows you to begin another measurement.
4. The Caliper measurement can display the following results:
 Distance
 Time
 Slope
HR: The HR measurement tool can provide the heart rate calculation.
To use the HR measurement tool:
1. Invoke the Measure Function on an M-mode strip.
2. If HR is not the default measurement, select HR from the touch screen.
3. Move the caliper to the first heart beat. Press the Set key to activate a second caliper.
4. Move the second caliper to position the next beat.
5. Press the Set key to complete or press the Update key to switch between moving the first
or second caliper.
6. The HR measurement can display the following result(the actual results shown will
depend on the preset):
 HR

7.3.2. Application Measurements and Calculations

Application Measurements have a pre-defined meaning and can be entered into a report. The
system supports the following application packages, each with its own set of measurements,
calculations and report:
 OB (including multiple fetuses)
 Abd
 Vascular
 Cardiac
 Gynecology
 Small Parts
 Urology
 Pediatric(including Neonatal)
Please see the Advanced User Manual for the complete content of each package.
When you select an application measurement from the touch screen it will automatically invoke
the type of measurement it needs. For example, if you select "BPD" from the OB application it
will automatically invoke a distance measurement. These measurements generally behave as
described above for generic measurements. However, a final press of the Set key after the
measurement is complete will enter the result into the report.
Some application measurements can have multiple variations. For example, in a twin OB exam
the OB measurements can be done on either fetus. As another example, in Vascular exams several
measurements can be done in a Proximal, Mid, or Distal location. When a measurement has
multiple variations you will see buttons on the touch screen that let you pick which variation you
are measuring.
- 87 -