CDN THD2FE Quick Start Manual

CDN THD2FE Quick Start Manual

Non-contact forehead thermometer
Hide thumbs Also See for THD2FE:

Advertisement

FOR MEASUREMENT OF HUMAN BODY
TEMPERATURE IN PEOPLE OF ALL AGES
• Response Time: Forehead mode: 1 second;
Surface mode: 0.6 second
• Forehead Measurement Distance: within 1.5 in/4 cm
• Surface mode: Distance:Spot = 1:1, emissivity: 0.95
• Accuracy: ±0.4°F/0.2°C
• Backlit in blue
• Memory recalls 25 readings
• One-button operation
• Battery status indication
• Temperature alert or silent
• On/Off Power button
• Auto-off after 1 minute
• ABS plastic
• Two 1.5V AAA IEC LR03 alkaline batteries, included
• FDA, RoHS, ISO 9001, ISO 13485, ASTM E1965-98,
EN ISO 80601-2-56
Note:
Remove label from display before initial use.
PO Box 10947 Portland, OR 97296-0947 Tel 800 338-5594 Fax 800 879-2364 Web CDNkitchen.com
NON-CONTACT FOREHEAD THERMOMETER
Note:
In the following instructions, names of the control
buttons are shown in CAPS. Function information that
appears on the display is shown in BOLD CAPS.
USAGE
Indications for Use
The Non-Contact Forehead Thermometer is an
infrared thermometer intended for the intermittent
measurement of human body temperature in people
of all ages.
Intended Operator:
At least 11 years old (5 years intensive reading
experience), no maximum.
Ambient Temperature
If there is any temperature difference between the
places where the device is stored and where it will
be used, allow the thermometer to adjust to the room
temperature where the subject is for at least
15 minutes before measurement.
Points of Attention
1. Forehead temperature is displayed in oral mode.
This mode converts the forehead temperature to
display its "oral-equivalent" value.
2. Before the measurement, the subject should stay
in a stable environment for 5 minutes and avoid
exercise or bath for 30 minutes.
3. Keep the forehead area clean and free from sweat,
cosmetics and scars while taking the temperature.
4. The "Clinical Bias" is -2.5 to -3.1°F (-1.4 to -1.7°C).
5. The "Limits of Agreement" is 0.98.
6. The "Repeatability" is 0.36°F (0.20°C)
BEFORE WE BEGIN
This thermometer has been designed for use in
professional healthcare facility environments and
home healthcare environments (as defined in 3.1 of
IEC 60601-1-11 Edition 2.0 2015-01 and Guidance
for Industry and Food and Drug Administration Staff:
Design Considerations for Devices Intended for
Home Use, Issued November 24, 2014). It is not
meant to replace a visit to the doctor. Compare the
measurement result to your regular
body temperature. Consult with doctor
if you have health concerns.
Forehead mode: 93.2 to 108°F/34 to 42.2°C
Surface mode: -7.6 to +176°F/-22 to +80°C
Model: THD2FE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the THD2FE and is the answer not in the manual?

Questions and answers

Summary of Contents for CDN THD2FE

  • Page 1 Model: THD2FE NON-CONTACT FOREHEAD THERMOMETER Forehead mode: 93.2 to 108°F/34 to 42.2°C Surface mode: -7.6 to +176°F/-22 to +80°C Note: In the following instructions, names of the control FOR MEASUREMENT OF HUMAN BODY buttons are shown in CAPS. Function information that appears on the display is shown in BOLD CAPS.
  • Page 2: Battery Installation

    A malfunction may occur if Silent the power is on when the battery is installed. If a i. When the THD2FE is on, press and hold malfunction occurs, restart the device. the POWER button for 3 seconds.
  • Page 3: Troubleshooting

    Select the target within (+80°C) specifications. If a malfunction • Do not tamper with the unit’s internal components. still exists, contact CDN for Forehead mode: Temperature Doing so will invalidate the warranty on the unit further assistance. taken is lower than 93.2°F (+34°C)
  • Page 4: Symbol Descriptions

    RF emissions Class B connected to the public low-voltage power supply field strength in the location in which the THD2FE is used exceeds the applicable RF compliance CISPR 11 network that supplies buildings used for domestic level above, the THD2FE should be observed to verify normal operation. If abnormal performance is purposes.
  • Page 5 TEST SPECIFICATIONS FOR ENCLOSURE PORT IMMUNITY TO RF WIRELESS COMMUNICATIONS EQUIPMENT The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE should assure that it is used in such an environment.