3. The donor/recipient block holder cylinder turns to the loading position with its appropriate
side. The ID of the corresponding tray is marked with green, also the slot is lit for easier
identification.
4. The door opener button turns green (button status is shown at the right side of the main
toolbar). The light pulsates for about 10 seconds and it is possible to open the door within
this time frame (countdown is displayed in place of the
5. Push the door opener button to open the door.
6. Press down the flap in front of the block slot.
7. Slide the block into position with the slanted side towards the back of the machine.
8. Release the flap when the block is in position.
9. Close the door.
10. The column turns into the next position, the height of the inserted block is measured (also,
the presence of the block is checked). These data are saved for later use (for drilling,
punching operations, and depth of insertion) and based on the measurements, the software
calculates the distance of the objective from the surface of the recipient block in order to
take an accurate photo of it.
11. Follow the steps 1-9 if you want to insert additional blocks into other trays. If the cylinder is
at the proper position, it is not necessary to the click the tray ID then the

4.4.4 Removing a block from the main unit

1. Recipient block positions are displayed in a double-column on the left side of the screen.
Donor block tray positions are identified by six buttons (A-F). The Change Blocks buttons are
responsible for rotation of the donor block holder cylinder.
2. Depending on which donor block tray is needed to be rotated to the block removal position,
click the tray ID (A-F). For the recipient blocks click the
specific block pairs.
3. A donor/recipient block holder column turns to the loading position with its appropriate side.
4. The door opener button turns green. The light pulsates for about 10 seconds and it is possible
to open the door within this time frame (countdown is displayed on the
5. Push the door opener button to open the door.
6. Press down the flap in front of the block slot.
7. Remove the block from its position and release the flap.
8. Close the door.
November 5, 2013 – Rev. 1