TMA Grand Master 2.0 – User's Guide
If TMA Grand Master cannot successfully identify the loaded PCR cassette, clean the
slot area and the cassette, then insert the cassette again. For cleaning blocks use
regular (37mm size) blocks only.
To avoid contamination, cleaning block are used before each PCR sampling. A cleaning block is a virgin
recipient block from which paraffin core is extracted and placed back with the puncher, hence
cleaning its surface.
Upon selecting the cleaning block on the user interface, an automatic layout is generated based on
the size of the puncher in use. If not every cleaning position had been used up during a previous
project session, you can use that block further on but make sure to manually mark the first free
position.
After the software is informed that recipient block loading is finished, it applies a layout consisting
spots of the same puncher diameter on the cleaning block.
Before starting a larger project it is advised to calculate the number of cores that will
be inserted to PCR tubes, so placing a sufficient number of cleaning blocks
accordingly.
The number of cleaning spots depending on the puncher diameter are the following:
Puncher diameter
0.6mm
1mm
1.5mm
2mm
November 5, 2013 – Rev. 1
Important!
Important!
Number of cleaning spots
252
144
72
50
3DHISTECH Ltd.
4 Operating the Main Unit
26(72)