Resetting the Mass Spectrometer
Resetting the Data System
Thermo Scientific
If communication between the mass spectrometer and data system computer is lost, it might
be necessary to reset the mass spectrometer using the System Reset button on the right-side
power panel. Pressing the System Reset button creates an interrupt in the embedded
computer. This causes the embedded computer to restart in a known (default) state. See
Figure 35
on
page 50
for the location of the System Reset button.
The procedure given here assumes that the mass spectrometer and data system computer are
both powered on and operational. If the mass spectrometer, data system computer, or both are
off, go to
"Starting Up the System after a Complete Shutdown"
To reset the mass spectrometer
Press the System Reset button located on the right-side power panel. Make sure the
Communication LED is extinguished before releasing the System Reset button. When you
press the Reset button, the following occurs:
• An interrupt on the embedded computer causes the CPU to reboot. All LEDs on the
front panel of the mass spectrometer are off except the Power LED.
• After several seconds, the Communication LED illuminates yellow to indicate that the
data system and the mass spectrometer are starting to establish a communication link.
• After several more seconds, the Communication LED illuminates green to indicate that
the data system and the mass spectrometer have established a communication link.
Software for the operation of the mass spectrometer is then transferred from the data
system to the mass spectrometer.
• After three minutes the software transfer is complete. The System LED illuminates either
green to indicate that the instrument is functional and the high voltages are on, or yellow
to indicate that the instrument is functional and is in standby.
There are two ways to reset the data system:
•
Resetting the Data System by Using the Windows Shutdown and Restart Procedure
•
Resetting the Data System by Turning the Personal Computer Off Then On
3
System Shutdown, Startup, and Reset
Resetting the Mass Spectrometer
on
page
54.
TSQ Series Hardware Manual
57