Input/Output Parameters - Pavone Systems MC 315 Series Technical Manual

Weight indicator wit serial outputs, fiber optics, in/out
Table of Contents

Advertisement

Fun. I n. 1
INPUT 1 FUNCTION
Selection of the function associated with input 1. [1301]
Selectable values:
Zero: Performs zero-point calibration. [0]
Tare: Performs auto-tare. [1]
Del.Tar: Deletes the tare. [2]
Send: On-demand data transmission. (**) [3]
Default: Zero
FUn. I n. 2
INPUT 2 FUNCTION
Selection of the function associated with input 2. [1302]
Selectable values:
Zero: Performs zero-point calibration. [0]
Tare: Performs auto-tare. [1]
Del.Tar: Deletes the tare. [2]
Send: On-demand data transmission. (**) [3]
Default: Zero
NodE 1
Set point 1 OPERATION MODE
Select 4 subsequent operation criteria of set point 1: [1303]
Compare with net weight, gross, weight, or peak. In the latter case, comparison occurs with the last
acquired peak value, even if the peak function is not enabled.
NET
The relay output is enabled in the Net Weight mode. [0]
GROSS
The relay output is enabled in the Gross Weight mode. [1]
Default: GROSS
Select the output status whether normally open or closed: [1304]
n.oPEn
Relay 1 is normally open. [0]
n.CLoSE
Relay 1 is normally closed. [1]
Default: n. oPEn.
Select whether positive or negative values should be compared: [1305]
PoSIt..
The output operates with positive weight. [0]
nEGAt.
The output operates with negative weight. [1]
ALL:
The output operates both with positive and with negative weight. [2]
Default:
PoSIt
Select whether stable only or both stable and unstable weight values should be compared: [1306]
norMAL
Output 1 is enabled with unstable weight. [0]
StAbLE
Output 1 is enabled with stable weight. [1]
Default:
norMAL

INPUT/OUTPUT PARAMETERS