Contents 1 Lojer Scandia SC330 Operating table ......................4 1.1 Content of the package ........................... 4 1.2 Description of parts ..........................5 1.3 Controls ..............................7 1.3.1 Hand control ............................. 7 1.3.2 Wireless (blue) Bluetooth hand control (Option) ................8 1.3.3 Foot control (Accessory) ........................
Page 3
Read these instructions carefully. Follow all warnings and instructions marked on the product Lojer Group is leading producers of medical furniture and physiotherapy equipment in Nordic Countries. We design and manufacture medical and treatment furniture to be used by health care professionals in various operating environments.
1 Lojer Scandia SC330 Operating table Lojer Scandia SC330 Operating table is designed to be used in surgical procedures for humans carried out in operating rooms by healthcare professionals. The operating table is designed to be primarily used with batteries. The operating table has various hydraulic adjustments which ensure diverse functions required in various operations.
1.2 Description of parts The parts of the Lojer Scandia SC330 operating tables are shown below (Figure 2). Figure 2: Lojer Scandia SC330 operating table in standard configuration 40 cm section (2 pcs.) 25 cm section (2 pcs.) Middle section...
The functions of the table can be operated also with a foot control which is available as an accessory (Figure 6). The functions of the foot control are programmed according to the customer’s order with the Lojer Service Software. 1-3 functions can be programmed. The symbols are identical with the hand control symbols. The foot control is connected to the table on the short section of the lower frame (see Figure 2 (10)).
3.5.7 Slide (Option) The horizontal slide of the table top is available as an options to Lojer Scandia operating tables. The table top will slide towards the long end of base frame by pressing and holding the hand control button (Figure 5 (12)).
In order to alter the speed, please contact Lojer Service. Figure 10: Using the driving wheel 3.6 Attaching/removing the table top sections The table top sections of the Lojer Scandia SC330 Operating table are easy to attach with a simple locking mechanism. ...
3.8 Charging the batteries The batteries of Lojer Scandia SC330 operating tables are connected in series and are designed for cyclic use. The lifetime of the batteries depend on the length of the charge/discharge periods. In order to ensure long lifetime of the batteries, charge the batteries as described below. The need for change depends on the internal resistance of the batteries (the internal resistance of a new battery is 12 mΩ...
Do not let hypochlorite solution to come into contact with metallic parts. Long-term contact with mattresses may damage them. If non-Lojer mattresses are used, refer to the table mattress manufacturer’s instructions on the use of disinfectants. Check mattresses regularly for damages and wearing. Replace mattress with damaged surface with new one immediately.
The most common failures and actions are presented on the table below (Table 4). The operating table can be operated in failure mode. The failure mode is presented by flashing of the indicator LEDs and repeated beep (see Table 2) when the table top is moved. Table 4: Troubleshooting of Lojer Scandia SC330 operatign table Symptom Cause...
Page 27
Contact Lojer Service in order to change the control. To order spare parts contact the Lojer Service. Before contacting, find out the following information from the type plate of the device: Name, model and the serial number of the device ...
LOJER shall assume the costs for the materials / spare parts and the labour costs as required, as well as the standard return transportation to the sender. In case of difficulty, details of local dealers are available online at www.lojer.com.