Lojer Afia 4040 Operating Instructions Manual

Lojer Afia 4040 Operating Instructions Manual

Examination tables

Advertisement

Quick Links

Afia 4040, 4050 and 4140
Examination tables
Operating instructions

Advertisement

Table of Contents
loading

Summary of Contents for Lojer Afia 4040

  • Page 1 Afia 4040, 4050 and 4140 Examination tables Operating instructions...
  • Page 3: Table Of Contents

    Operating instructions 4040, 4050 and 4140 examination tables/Version 1.15 Contents Lojer Afia 4040, 4140 and 4050 examination tables ..............1 Parts of the examination tables....................... 1 Afia 4040 ............................1 Afia 4050 ............................2 Afia 4140 ............................3 Hand control ............................ 4 Options and accessories for examination tables ..............
  • Page 4 Technical data ..........................20 Applied standards ........................21 Troubleshooting ..........................21 Spare parts ............................ 22 Lojer Warranty ..........................23 11.1 International warranty conditions ................... 24 11.2 Proceeding in a warranty case ....................24 Contact information ........................25 Copyright © Lojer Oy, 2007...
  • Page 5: Lojer Afia 4040, 4140 And 4050 Examination Tables

    4040, 4050 and 4140 examination tables/Version 1.15 Lojer Afia 4040, 4140 and 4050 examination tables Lojer Afia 4040 examination and treatment table, 4050 gynaecological general examination table and 4140 general examination table are intended for use in examinations and treatments by medical staff in hospitals and clinics.
  • Page 6: Afia 4050

    Operating instructions 4040, 4050 and 4140 examination tables/Version 1.15 Figure 1. Afia 4040 examination and treatment table Afia 4050 Hand control Paper roll holder Mounting adapter Infusion pole (accessory) Arm support plate (accessory) Anaesthesia screen (accessory) Backrest Gynaecolocical backrest Adjusting lever for gynaecological backrest...
  • Page 7: Afia 4140

    Foot control Central locking pedal Figure 2. Afia 4050 gynaecological general examination table Afia 4140 Hand control Paper roll holder Backrest Heel support Bowl Foot control Central locking pedal Figure 3. Afia 4140 general examination table Copyright © Lojer Oy, 2007...
  • Page 8: Hand Control

    Examination tables are available with following options: Separately lockable wheels Trendelenburg function Special upholstery Following accessories are available for examination tables: Side rails Mounting adapter D20 for accessories Adjustable D20 fastener Infusion pole Arm support plate Handrails Leg supports Neck cushion Copyright © Lojer Oy, 2007...
  • Page 9: Symbols Used On The Examination Tables

    4040, 4050 and 4140 examination tables/Version 1.15 Many accessories for operation tables with standard attachment to side rails (More 12 Contact information detailed information from Lojer sales, see section See the picture of Afia 4050 examination table for more detailed information about 2.2 Afia 4050...
  • Page 10: Commissioning

    NOTE! A 5 mm Allen key and a 13 mm spanner are required for mounting. Maximum three side rails can be mounted on both sides of the examination table . Remove plastic plugs (3 pcs) from the side of the examination table. Insert bushings (3 st) into the holes. Copyright © Lojer Oy, 2007...
  • Page 11: Using The Examination Table

    Maximum movement and adjustment ranges of the table top sections are illustrated below. Table top movements 4040 and 4140 Table top movements 4050 Height adjustment 50 -95 cm Figure 6. Maximum movement and adjustment ranges of the examination tables Copyright © Lojer Oy, 2007...
  • Page 12: Locking The Wheels

    (risk of squeezing). 5.3.1 Height adjustment with hand control You can raise or lower the examination table by pressing the hand control buttons. Raising the table Lowering the table Figure 8. Height adjustment with hand control Copyright © Lojer Oy, 2007...
  • Page 13: Height Adjustment With Foot Control

    On the side of the examination table the table top can be raised by moving the foot control towards the front end of the table and lowered by moving the foot control towards the rear end of the table. Copyright © Lojer Oy, 2007...
  • Page 14: Adjusting The Backrest

    The backrest locks into place when the handle is released. Gynaecological backrest can be raised to 45° angle. The gynaecological backrest is lowered by turning the handle and pressing the backrest to desired position. Copyright © Lojer Oy, 2007...
  • Page 15: Adjusting The Seat (4050)

    Lower the heel support and pull or push it so that it is locked in a locking hole. There are five locking holes in the heel support arms. Check always that the heel support is locked into its place. Copyright © Lojer Oy, 2007...
  • Page 16: Swiveling In/Out And Removing The Bowl (4140 And 4050)

    The paper roll holder is located under the front end of the examination table. The paper roll holder axle is locked into its place. Suitable paper rolls are available from Lojer sales. 1. Remove the paper roll axle by releasing the locks (2) with fingers.
  • Page 17: Trendelenburg Function

    Set the mounting adapter on the side rail. Push the accessory shaft/pole through the mounting adapter hole. Fasten the mounting adapter with accessory by tightening the adjuster screw. Copyright © Lojer Oy, 2007...
  • Page 18: Using The Arm Support Plate

    Loosen the arm support by opening the stem mounting adapter screw and cushion locking screw. Move the arm support plate to the desired position. Tighten the arm support into its position with the mounting adapter screw and cushion locking screw. Copyright © Lojer Oy, 2007...
  • Page 19: Raising And Lowering The Hand Rail

    Attach the neck cushion (accessory) to the armrest with the loop on the back of the cushion. Figure 20. Neck cushion 5.15 Leg supports (4050 and 4140)) Leg supports are attached on the rear end on the examination table. Copyright © Lojer Oy, 2007...
  • Page 20: Height And Width Adjustment Of The Leg Supports

    Push the infusion pole through the mounting adapter hole. Tighten the infusion pole into its position with the mounting adapter screw. Height of the infusion pole can be adjusted with a screw on the upper part of the pole. Copyright © Lojer Oy, 2007...
  • Page 21 Operating instructions 4040, 4050 and 4140 examination tables/Version 1.15 Infusion pole Mounting adapter Figure 22. Infusion pole Copyright © Lojer Oy, 2007...
  • Page 22: Mounting Of Anaesthesia Screen

    The surface of skai fashion upholstery and fashion materials is not resistant to solvents, chlorides, polishing agents, washing/polishing agents, and aerosol sprays. Please note that colourings by jeans or other textiles are excluded from any manufacturer‘s guarantee. Copyright © Lojer Oy, 2007...
  • Page 23: Disinfection

    Disinfection You can disinfect the parts of the examination table for example with ERISAN Des Contact information. disinfection agent (available from Lojer sales, see Section 12 the package for dilution instructions) NOTE! ERISAN Des disinfection agent is not suitable for synthetic leather upholsteries.
  • Page 24: Technical Data

    230 W Operating time ratio 2 min. / 18 min. Enclosure class IPX4 Protection class Class II, type B Weight ~110 kg (without options) Maximum load (SWL) 200 kg Width 65 cm Length 193 cm Copyright © Lojer Oy, 2007...
  • Page 25: Applied Standards

    SFS-EN 349 SFS-EN 60601-1 IEC 601-1:1988 + A1:1991 + A2:1995. Troubleshooting NOTE! The warranty is not valid if a motor or the control unit is opened by someone else than a Lojer authorized service person. Symptom Trouble Action The motor The wiring is not properly Check the wiring connections.
  • Page 26: Spare Parts

    Test the suspected control with another similar examination table and replace if necessary. For changing a motor, hand or foot control or control unit contact Lojer service (see Contact information Section 12 ). Please always have the following information ready...
  • Page 27: Lojer Warranty

    The Lojer product is warranted for the period of twenty-four (24) months from the original date of purchase, against defective materials and workmanship. In the event that warranty service is required, you should return the product to the dealer from whom it was purchased.
  • Page 28: International Warranty Conditions

    3. In case of warranty repairs, they must be carried out by Authorised Lojer Dealers. No re-imbursement will be made for repairs carried out by non Lojer representative and, any such repair work and damage to the products caused by such repair work will not be covered by this warranty.
  • Page 29: Contact Information

    The model designation, the article code and the serial number of the product The date of purchase A brief description of the problem. Contact information Manufacturer Lojer Oy Putajantie 42, FI-38201 Sastamala, FINLAND Tel. +35810 830 6700, Fax. +35810 830 6702 Email: firstname.lastname@lojer.com www.lojer.com Copyright © Lojer Oy, 2007...

This manual is also suitable for:

Afia 4140Afia 4050

Table of Contents