Lojer Delta Series Instructions Of Use

Treatment tables

Advertisement

Quick Links

Lojer Delta
treatment tables
Instructions of use
18.2.2013

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Delta Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Lojer Delta Series

  • Page 1 Lojer Delta treatment tables Instructions of use 18.2.2013...
  • Page 2: Table Of Contents

    Contents Lojer Delta- treatment tables ............................. 3 Desciption of parts ................................3 Options and accessories ..............................4 Symbols used on the device .............................. 5 Introduction ..................................5 Inspection upon delivery ..............................5 Before use ..................................6 Using the treatment table ..............................6 Maximum inclination .................................
  • Page 3: Lojer Delta- Treatment Tables

    Lojer Delta- treatment tables Lojer Delta treatment tables are intended for professionals to be used for massage and wide range of physiotherapy. The tables are designed for a treatment given on the lying surface. Delta-tables can also be used as an examination table.
  • Page 4: Options And Accessories

    Castors Head section, head section adjustment lever Lying surface Central locking pedal Foot control Options and accessories Options and accessories available for Lojer Delta treatment tables: Place free height adjustment bar Castors Double hinged head section Handswitch Foot pedal Frame for foot pedal (Standard and Professional) Copyright ©...
  • Page 5: Symbols Used On The Device

    Ensure that the delivery contains all the parts detailed in the delivery note. If there is anything missing from the delivery consignment, please contact the supplier immediately. The device can be stored at a temperature of -5...+60 °C. The permitted humidity is 30...75 %. Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...
  • Page 6: Before Use

    The height adjustment range of the Delta treatment tables: Standard; 1MD4 57-87 cm, 1MD7L and 2MD602 58-89 cm Professional; 1MD4 47-98 cm, 1MD7L and 2MD602 48-100 cm. The foot section adjustment range to 75 ° angle. Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...
  • Page 7: Central Locking Of The Castors (Optional)

    WARNING! Children or people with no experience of the device must not use the device. For safety reasons use the safety switch or unplug the power cord when the device is left without supervision. Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...
  • Page 8: Adjusting The Middle Section (Drainage)

    Lift the adjustment lever (1) (Figure 7) and lift/push the head section to the desired position. Figure 7: Adjusting the head section Max. load of the head section is 50 kg. Do not lift the table on the head section. Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...
  • Page 9: Adjusting The Arm Rests

    Double hinged face section (accessory) The counter adjustment can be adjusted by freeing the adjustment lever (1) while lifting/pressing the head section (Figure 10). Figure 10: Adjusting the double hinged head section Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...
  • Page 10: Cleaning And Disinfecting

    Disinfecting wears out the surfaces. After disinfecting clean the surfaces with clean, damp cloth. Dilute the disinfectant according to the manufacturer’s instructions. Upholstery Skai-material: PROPERTIES: LONG SERVICE LIFE FLAME RETARDANT AGREEABLE IN USE RESISTANT TO DISINFECTANTS ABRASION RESISTANT TEAR-RESISTAMT Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...
  • Page 11: Maintenance

    Recommendable cleaning agent: Lojer Desiplint (1:10), which is effective against bacteria without drying the upholstery material. Repeat the cleaning procedure if necessary. (Composition of Lojer Desiplint: Chlorhexidine-digluconate 0,1 – 0,2 %, water 99,8 Use the recommended disinfectants (get matrix from service@lojer.com). Lojer cannot be held responsible for the effects of other solutions than those mentioned in the matrix.
  • Page 12: Annual Measures

    The actuator might stop working. Contact the service. out or overloaded. In order to change the actuators, control or control box and ordering spare parts contact the Lojer service. Before contacting the service, find out the following information from the type plate of the device.
  • Page 13 RESULSTS identification of the testing organization, name of the person who performed the tests, identification of the equipment, details of the tests, date and the result of the functional tests and measurements. Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...
  • Page 14: Technical Information

    Storage temperature -5…+60°C, humidity 30…75 % Operating temperature +10…+40 °C, humidity 30…75% Figure 11: Type plate Circuit diagrams Circuit diagram is shown below (Figure 12). Figure 12: Circuit diagram of Delta tables Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...
  • Page 15: Standards

    HOUSE HOLD WASTE: plastic, upholsteries and other materials which cannot be separated further. Contact your local disposal authority for more details of how to recycle. Follow the instructions given in local collection points. Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...
  • Page 16: Contact Information

    FI-38201 Sastamala Tel.. +35810 830 6700 Fax. +35810 830 6702 Email: firstname.lastname@lojer.com info@lojer.com www.lojer.com Your local Lojer dealer, see www.lojer.com/distributors Model: ________________________________________________________ Serial number: __________________________________________________ Date of purchase: ________________________________________________ Your local Lojer dealer:____________________________________________ Copyright © Lojer Oy, 2013 02/2013 Rev. 1.0...