6 Troubleshooting and Maintenance

6.1 Safety Information

Check cables, plugs and connections before operation, and if defective or cannot be replaced, contact
3DHISTECH Service for support.
Check whether the device is compatible with your local line voltage. It can operate at line voltages
100V to 240V and 50/60 Hz.
Only plug the device in a socket with protective grounding (earth) contact.
Do not operate the device in explosive environment.
Disconnect the device from the power outlet before opening it or replacing a part.
The product is not equipped with protection against hazardous materials (infectious, toxic,
radioactive, corrosive substances or other effects damaging to health). Decontaminate infectious
materials after use.
Dust, specks and humidity may decrease or harm the operating capability of the device. Protect it
with dust cover when not in use. Switch the power off before covering it.
Defective devices do not classify as domestic waste. Dispose of the device in compliance with the
currently valid legal requirements.
Slides that are not used any longer do not classify as domestic waste and must be treated as
hazardous materials and disposed in compliance with the currently valid legal requirements.

6.1.1 Cleaning the device

Before cleaning, switch off the device and pull the plug out.
Do not let liquid penetrate into the system.
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth with water.
To clean greasy surfaces, it is recommended to apply Isopropanol on a piece of lint-free cotton cloth
and gently rub the surface.
March 19, 2019 – Rev. 2
3DHISTECH Ltd.
6 Troubleshooting and Maintenance