Label Area – Modify slide label settings if necessary. Turn on Rotate 180° to rotate label by
180 degrees.
Color Correction – Different color profiles can be assigned to each slide.
ICC Profile – Select an applicable preset color profile to be used for Brightfield scanning
from the drop-down menu. Settings according to the profile has effect on the color set,
color temperature, and gamma value of the scanned image. Preset profiles are the
following:
Color Scheme
White Point
Gamma
March 19, 2019 - Rev. 2
Calibrated Linear: Color scheme is defined by calibration (Gamma value is not
included)
sRGB standard: Color scheme is defined by calibration (standard Gamma value -2.2-
is applied)
sRGB standard Microscope Like: Color scheme is defined by calibration (standard
Gamma value / White Point: D50)
Custom: parameters are freely customizable as follows:
Calibrated: Scheme is defined by the color calibration slide
Raw: the color information of the recorded raw image is applied
D65: color temperature according to a standard illuminant D65 white point value is
applied (6500K / daylight)
D50: color temperature according to a standard illuminant D50 white point value is
applied (5000K / horizontal daylight)
Standard: a value of 2.2 is used
Medium: a value of 1.8 is used
Off: a value of 1.0 is used
3DHISTECH Ltd.
4 Software and Work-flow
42(78)