Routine Schedule - Carlisle Ransburg RansFlex RFXi Series Service Manual

Applicators for in-direct charge waterbased applicators
Table of Contents

Advertisement

EN
CAUTION
!
† Nozzles from previous Ransburg design are not
compatable with the Ransflex design. Use of these
nozzles could cause equipment malfunction and
possible damage.
WA R N I N G
!
† The flash point of the cleaning solvent shall be at
least 15° C (27° F) above the ambient temperature.
Otherwise, the cleaning process must be carried out in
an area with forced air ventilation. It is the end users
responsibility to insure this condition is met.
All repairs should be made on a clean, flat surface. If a
vise is used to hold parts during service or repair, DO NOT
clamp onto plastic parts and always pad the vise jaws!
The following parts should be thoroughly packed with
dielectric grease (LSCH0009-00) leaving NO air space or
voids when assembling:
• All O-Rings (PTFE o-rings do not need lubrication)
• Needle Shaft Assembly
• Packing Tube
• Cascade and Barrel
Equipment Required
• Special Multi-Purpose Wrench (80353-00)*
• Hex Driver (79862-02)*
• Dielectric Grease (LSCH0009-00)*
• 10 mm Wrench
• 15 mm Wrench
• Spanner
* Supplied with applicator

ROUTINE SCHEDULE

Follow these maintenance steps to extend the life of the
applicator and ensure efficient operation:
Several Times Daily
Inspect the air cap for paint accumulation. Clean as
frequently as necessary with a soft bristled brush and a
suitable solvent.
AH-17-02-R1 (10/2019)
† NEVER remove the fluid nozzle assembly while
paint is in the applicator or paint may enter into the
air passages. Clogged or restricted air passages will
cause poor atomization and/or electrical shorting. Air
passages that are clogged with conductive material can
lead to excessive current output levels and consequent
low operating voltage or long-term electrical damage.
The applicator barrel MUST be tilted front down to
remove the fluid nozzle. Failure to do so may allow paint
to enter the air passages, thereby reducing airflow and
damaging the applicator barrel/cascade. Applicators
may be flushed in lieu of tilting. However, they must
be either flushed or tilted down during nozzle removal!
Cleaning Flushing
1. Reference page 24 for flushing procedure. Flushing
should be performed at the end of daily use or prior to
any maintenance.
2. Applicator exterior cleaning at the end of each shift,
wipe the outside of the applicator with a solvent soaked
rag pointing the applicator nose down.(Use non-polar
solvents)
26 / 55
MAINTENANCE
CAUTION
!
www.carlisleft.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ransburg ransflex rfxi 81465

Table of Contents