Download Print this page
MadgeTech UltraShock Product User Manual

MadgeTech UltraShock Product User Manual

Temperature, humidity, pressure and tri-axial shock data logger

Advertisement

Quick Links

ULTRASHOCK
Temperature, Humidity, Pressure
and Tri-Axial Shock Data Logger
PRODUCT
USER GUIDE
To view the full MadgeTech product line, visit our
website at madgetech.com.

TABLE OF CONTENTS

Advertisement

loading

Summary of Contents for MadgeTech UltraShock

  • Page 1: Table Of Contents

    Tri-Axial Shock Data Logger PRODUCT TABLE OF CONTENTS USER GUIDE 2 Product Overview 4 Software Installation Guide To view the full MadgeTech product line, visit our website at madgetech.com. 5 Device Operation 8 Product Maintenance 9 Specifications 10 Dimensional Drawings...
  • Page 2: Product Overview

    • Adhesive joint material should be as stiff as possible (high durometer) The stiffness of the mounting material will affect the transfer of shock or vibration to the UltraShock. A soft material will have a tendency to reduce the transfer of high frequencies while a hard material would, in varying degrees, come closer to the fidelity of signal that a bolted down mount would provide.
  • Page 3 UltraShock PRODUCT OVERVIEW LED Indicator Functions The red and green LEDs on the front of the UltraShock will flash or maintain a static condition to represent operating functions as shown Red LED Indicator in the table below. The power up indication is only presented if the battery pack is separated Green LED from the front face and replaced.
  • Page 4: Software Installation Guide

    2. Locate and unzip the downloaded file (typically you can do this by right clicking on the file and selecting Extract). 3. Open the MTInstaller.exe file. 4. You will be prompted to select a language, then follow the instructions provided in the MadgeTech 4 Setup Wizard to finish the MadgeTech 4 Software installation.
  • Page 5: Device Operation

    UltraShock DEVICE OPERATION Connecting and Starting the Data Logger Enabling and Disabling of Channels 1. Once the software is installed and running, plug the 1. Connect the logger to the interface cable. interface cable into the data logger. 2. Highlight the data logger in the Connected Devices list.
  • Page 6 Logging Mode - Peak vs. Instantaneous When Custom Start is selected in the MadgeTech 4 Software, the user is presented with a range of logging time intervals or actual recording rates (sampling rates). For all but the two highest recording rates, the user will also have a choice of Peak vs.
  • Page 7 The trigger sample count should be chosen carefully because it may gather data points for which we have no The UltraShock allows the user to enter a shock value interest. For instance, if 100 were entered and the logging which will trigger the storage of data. In general, consider...
  • Page 8: Product Maintenance

    Inspection The UltraShock battery does not require any periodic The UltraShock may be inspected visually for any signs of maintenance. It should be charged as needed. If the damage, especially to the gray gasket or to the USB jack. If...
  • Page 9: Specifications

    TEMPERATURES ABOVE THE MAXIMUM OPERATING TEMPERATURE OF THE PRODUCT. DISPOSE OF PROPERLY. CHARGE ONLY WITH THE PROVIDED MADGETECH CHARGER, OR FROM OTHER USB POWER SOURCE VIA THE MADGETECH PROVIDED CABLE. Specifications subject to change. See MadgeTech’s terms and conditions at madgetech.com.
  • Page 10: Dimensional Drawings

    TYPE C GASKET USB CONNECTOR 1.34 [34.0 mm] 1.055 [26.8 mm] 1.179 [29.9 mm] 0.885 [22.5 mm] BATTERY PACK FRONT VIEW BACK VIEW (BOTTOM) 6 Warner Road, Warner, NH 03278 (603) 456-2011 • info@madgetech.com • madgetech.com DOC-1363036-00 REV 4 2020.08.12...