Chapter 3
VXI-MXI-2 Configuration and Installation
Do not install a device configured for Slot 0 into another slot without first
Caution
reconfiguring it to either Non-Slot 0 or automatic configuration. Neglecting to do this
could damage the device, the VXIbus backplane, or both.
PCI-MXI-2 for Linux
a. Automatic Slot 0 Detection (Default)
Figure 3-3. VXIbus Slot Configuration
When the VXI-MXI-2 is installed in Slot 0, it becomes the VMEbus
System Controller. In this role, it has VMEbus Data Transfer Bus Arbiter
circuitry that accepts bus requests on all four VMEbus request levels,
prioritizes the requests, and grants the bus to the highest priority requester.
As VMEbus System Controller, the VXI-MXI-2 also drives the 16 MHz
VMEbus system clock by an onboard 16 MHz oscillator.
W2
Slot 0
Auto
Nonslot 0
W2
Slot 0
Auto
Nonslot 0
b. Manual Slot 0 Configuration
W2
Slot 0
Auto
Nonslot 0
c. Manual Nonslot 0 Configuration
3-6