Vxi/Vme-Mxi-2 - National Instruments MXI Series Getting Started

Hide thumbs Also See for MXI Series:
Table of Contents

Advertisement

Chapter 1
Introduction and Quick Start

VXI/VME-MXI-2

PCI-MXI-2 for Linux
This section summarizes the hardware and software default settings for the
VXI-MXI-2 and VME-MXI-2.
Table 1-6. VXI-MXI-2 Hardware Default Settings
Hardware Component
Logical Address (U43)
VXIbus Slot 0/Non-Slot 0 (W2)
VXIbus Local Bus (S8, S9)
VXIbus CLK10 Routing (W3)
External Trigger Termination (S2)
SMB CLK10 Direction (S3)
SMB CLK10 Termination (S4)
Polarity of External SMB CLK10
(S5)
MXIbus CLK10 Signal (S7)
MXIbus Termination
(U35 switches 1 and 2)
Configuration EEPROM
(U35 switches 3 and 4)
DRAM SIMMs Installed
SIMM Size Configuration (S6)
1
Automatic detection
Both OFF: Single VXI-MXI-2
From onboard oscillator
OFF: Unterminated
OUT: Drive CLK10 signal
Ignored; effective only when S3
is set to IN.
Inverted
Receive CLK10 from MXIbus
Automatic MXIbus termination:
switch 2 set to YES; switch 1
ignored.
User-modifiable; factory settings
protected: both switches set to
NO.
Per customer order
OFF if SIMMS are 4 M × 32
or larger; ON if smaller than
4 M × 32.
1-12
Default Setting
ni.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pci-mxi-2Vxi-mxi-2Vme-mxi-2

Table of Contents