Installation

SW 1 (Unit ID) Selector

Use the SW1 selector to choose a different unit ID. For the location of the selector, refer to the
The unit ID identifies the sample changer to software packages that can issue GSIOC commands to the sample changer.
The unit ID is set to 10. There is no need to change this number unless it is the same as that assigned to another Gilson instrument
that is also connected along the GSIOC.
To change the unit ID:
1
Gently insert a small flat-blade screwdriver into the SW1 selector on the rear panel and turn it.
2
Align the white dot with one of the indicated numbers. The unit ID is 10 plus the selected number.

SW2 (Baud Rate/Configuration) Selector

Use the SW2 selector to choose a different baud rate. For the location of the selector, refer to the
page 2-12.
As a default, the baud rate selector is set to 0 for identifying a baud rate of 19200 for computer control or for identifying an
external clock for GSIOC control.
For additional baud rate selections, refer to the rear panel label on the sample changer.
To change the baud rate/configuration setting:
1
Gently insert a small flat-blade screwdriver into the SW2 selector on the rear panel and turn it.
2
Align the white dot with one of the indicated numbers.

Fuse Installation

1
Locate the fuse drawer and two of the supplied fuses.
2
Insert the fuses in the fuse drawer.
3
Insert the fuse drawer into its receptacle on the rear panel.

Power Connection

Locate the appropriate power cord for your line voltage.
Use the power cord to connect the sample changer to a power source.
223 Sample Changer User's Guide
Rear Panel Diagram
on page 2-12.
Rear Panel Diagram
Fuse