Setting Of The Products - Pfeuffer HE lite Operating Instructions Manual

Grain moisture meter
Hide thumbs Also See for HE lite:
Table of Contents

Advertisement

Operating instructions HE lite
7.4

Setting of the products

After entering the code (see chapter 7.3)
the product list is indicated on the
display in the same way as for the
product selection during the measure-
ment:
→ Barley
Oats
Set
Select the product to be adjusted by
means of the arrow keys and confirm it
by pressing
.
On the display the calibration factors
offset (constant factor, factory setting
»0.00«) and gain (proportional factor,
factory setting »1.000« are indicated.
Barley
→ Offset:
Gain
The readings can be increased or
decreased by means of the arrow keys.

© 2017 – Pfeuffer GmbH
0.0
: 1.000
+
Set
(Example)
→ Offset:
Notice:
The calibration of the instrument must
be prepared carefully. Pay attention to a
homogeneous sample material without
dirt and green grain portions. Do not
change
the
measurement only.
Carry out at least three measurements
with the HE lite and calculate the
average.
Calculate the average deviation to the
comparative instrument.
Example:
You obtained the values 15.7 %, 16.0 %
and 15.9 % with the same sample. The
average reading is 15.9 %.
The grain dealer measures 15.4 % with
his instrument, i. e. 0.5 % less.
This means that the reading of the HE lite
is 0.5 % would be too high.
In this case the parameter offset must be
set to »-0.5« so that also the HE lite
indicates an average reading of 15.4 %.
Control the modification by means of
your original sample.
Change to the next factor »GAIN« by
pressing
Revision 9
Barley
-0.5
Gain
: 1.000
+
Set
calibration
after
.
Page 15 of 28
one

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1170 0071

Table of Contents