Calibration procedure
To carry on the recalibration the following instruments are needed:
- Known concentration gas cylinder suitable for the sensor, complete with a pressure regulator
- Flow meter
- Hose with Tee fitting to connect the cylinder to the flow meter and to the instrument
In the following is described a recalibration example for the CO sensor.
1. Start the instrument
Keep
pressed for
a few
seconds
ATTENTION
Make sure autozero is performed in clean air and completes properly.
Do not connect the gas probe to the instrument.
Check the battery charge level or connect the power adapter to avoid data loss
during recalibration.
2. Once autozero is completed press the
07/08/14
10:00
Diagnostic
Sensors
Gas probe
Hardware
Pump
On site cal.
E4500-N
Serial number: 1000
Firmware version: 1.00
Password
On site calibration
OK
0 0 0 0
OK
07/08/14
Combustion analysis
O
2
%
CO
2
%
Ex. Air
1.25
T flue
190.1
°C
T air
15.4
°C
∆T
74.7
°C
Loss tot
%
Eff. tot
91.4
%
key and select the diagnostic icon.
07/08/14
10:00
Enter the
recalibration menu
password 1111.
54
10:00
Combustion analysis
O
4.2
2
%
CO
9.3
2
%
Ex. Air
T flue
°C
T air
°C
∆T
°C
Loss tot
8.6
%
Eff. tot
%
Password
On site calibration
1 1 1 1
OK
K000000000EJ 023826A0 100714
07/08/14
10:00
4.2
9.3
1.25
190.1
15.4
74.7
8.6
91.4
07/08/14
10:00
OK