Calibration procedure
To carry on the recalibration the following instruments are needed:
- Known concentration gas cylinder suitable for the sensor, complete with a pressure regulator
- Flow meter
- Hose with Tee fitting to connect the cylinder to the flow meter and to the instrument
In the following is described a recalibration example for the CO sensor.
1. Start the instrument
Keep
pressed for
a few
seconds
ATTENTION
Make sure autozero is performed in clean air and completes properly.
•
Do not connect the gas probe to the instrument.
•
Check the battery charge level or connect the power adapter to avoid data loss
•
during recalibration.
2. Once autozero is completed press the
07/08/14
10:00
Diagnostic
Sensors
Gas probe
Hardware
Pump
On site cal.
E4500-N
Serial number: 1000
Firmware version: 1.00
Password
On site calibration
OK
0 0 0 0
OK
07/08/14
Combustion analysis
O
2
%
CO
2
%
Ex. Air
1.25
T flue
190.1
°C
T air
15.4
°C
∆T
74.7
°C
Loss tot
%
Eff. tot
91.4
%
key and select the diagnostic icon.
07/08/14
10:00
Enter the
recalibration menu
password 1111.