USER MANUAL
E500 Combustion analyzer
www.kimo-instruments.com
E Instruments, a member of Sauermann Group

Advertisement

Table of Contents
loading

Summary of Contents for E Instruments E500

  • Page 1 USER MANUAL E500 Combustion analyzer www.kimo-instruments.com E Instruments, a member of Sauermann Group...
  • Page 3: Table Of Contents

    TABLE OF CONTENTS IMPORTANT INFORMATION Information about this manual Danger levels and other symbols SAFETY Safety check Intended purpose Improper use of the product Precautions for the usage of the Li-Ion battery package WORKING PRINCIPLE General overview of the analyzer DESCRIPTION OF THE PRODUCT Working principle Measurement cells...
  • Page 4 TABLE OF CONTENTS MENU Setting Menu Menu→Measurements 9.2.1 Menu→Measurements→Draft 9.2.2 Menu→Measurements→CO air 9.2.3 Menu→Measurements→Pressure Menu→Configuration 9.4.1 Menu→Configuration→Analysis 9.4.1.1 Menu→Configuration→Analysis→Fuel 9.4.1.2 Menu→Configuration→Analysis→Condensation 9.4.1.3 Menu→Configuration→Analysis→O2 Reference 9.4.1.4 Menu→Configuration→Analysis→Measure units 9.4.1.5 Menu→Configuration→Analysis→Autozero 9.4.1.6 Menu→Configuration→Analysis→Air temp 9.4.1.7 Menu→Configuration→Analysis→NOx/NO Ratio 9.4.2 Menu→Configuration→Instrument 9.4.2.1 Menu→Configuration→Instrument→Display 9.4.2.2 Menu→Configuration→Instrument→On site calib.
  • Page 5 TABLE OF CONTENTS ANNEX A - Analysis report examples ANNEX B - Heading of the printed report with the PC Software “Easy2print” ANNEX C - Optional Measurements list / Measurement units matching → abbreviations ANNEX D - Coefficients of the fuels and Formulas ANNEX E - Declaration of Conformity...
  • Page 6: Important Information

    1.1 Information about this manual  This manual describes the operation and the characteristics and the maintenance of the Combustion Analyser model E500.  Read this operation and maintenance manual before using the device. The operator must be familiar with the manual and follow the instructions carefully.
  • Page 7: Safety

    This chapter describes the areas of application for which the E500 is intended. Using the E500 in other application areas is on the risk of the operator and the manufacturer assumes no responsibility and liability for loss, damage or costs which could be a result. It is mandatory to read and pay attention to the operating/maintenance manual.
  • Page 8: Working Principle

    - Generation and visualization of a QR code with the purpose of downloading the data of the acquired measures, having installed the App E instruments “E-Instruments QR Code App” which can be downloaded form the AppStore (if the version of the instrument foresees it).
  • Page 9: Description Of The Product

    Exhausted cells can be easily replaced by the user without depriving himself of the instrument and without complicated calibration procedures with certified mixtures as they are pre-calibrated before being supplied. E instruments certifies the accuracy of the measurements only upon a calibration certificate issued by its laboratory or other approved laboratory.
  • Page 10: Bluetooth® Connection

    4.9 IR connection The E500 analyzer is internally equipped with an infrared light interface which uses the HP-IR protocol, which allows the communication with a remote IR printer. 4.10 Software and available applications Easy2print PC Software for systems provided with Windows XP or later versions, downloadable from the web site www.e-...
  • Page 11: Components Description

    5.0 COMPONENTS DESCRIPTION 5.1 Instrument interface DESCRIPTION: Polyester keyboard with preformed keys and main command functions: KEYS FUNCTION Activates the context keys shown on the display. - Turns on and off the instrument. - If pressed briefly, accesses the instrument menu. - If pressed for at least 2 seconds, turns off the instrument.
  • Page 12 Display LCD backlit Display, White/Black 128 x 128 pixel with white LEDs. CAUTION: If the instrument is exposed to extremely high or extremely low temperatures, the quality of the display may be temporarily impaired. Display appearance may be improved by acting on the contrast key. Date, time and battery status.
  • Page 13: Technical Specifications

    6.0 TECHNICAL SPECIFICATIONS 6.1 Technical specifications Power: Li-Ion battery pack with internal protection circuit. Average life of the battery package: 500 empty / full charge cycles. Battery charger: External 5Vdc 2A battery charger with female A-type USB connector + connection to the device with the same serial communication cable supplied.
  • Page 14: Measurement And Accuracy Ranges

    6.2 Measurement and Accuracy Ranges RESPONSE MEASUREMENT SENSOR RANGE RESOLUTION ACCURACY TIME T90 Electrochemical 0 .. 25.0% vol 0.1% vol ±0.2% vol <20 sec. sensor ±20 ppm 0 .. 400 ppm Electrochemical 0 .. 4000 ppm 1 ppm <30 sec. with NOx filter sensor ±5% measured value...
  • Page 15: Using The Flue Gas Analyzer

    Remove the instrument from its packing and check it for damage. Make sure that the content corresponds to the items ordered. If signs of tampering or damage are noticed, notify the E INSTRUMENTS service center or agent immediately and keep the original packing. A label at the rear of the analyzer bears the serial number. This serial number should always be stated when requesting technical assistance, spare parts or clarification on the product or its use.
  • Page 16: Use With External Power Pack

    The instrument offers the possibility to generate and display a QR code with the purpose to download the data of the acquired measures, activating the interactive function “Print” visible on the display in the menu configuration, having installed the “E Instruments QR Code App” downloadable from the AppStore. Minimum requirements for the App installation “E Instruments QR Code App”...
  • Page 17: Connection Diagram

    7.5 Connection diagram AA AL05 AACTA03A AA SF7-A AAC KP02...
  • Page 18: Flue Gas Analysis

    BEING USED FOR PRIMARY AIR. ATMOSPHERIC BOILERS / FURNACES: UPON COMPLETION OF THE AUTOZERO IN FRESH OUTDOOR AIR, MOVE THE E500 INDOORS TO THE EXACT LOCATION OF THE APPLIANCE AND ONCE THE TEMPERATURE STABILIZES TO THE INDOOR TEMPERATURE, PRESS THE “KEEP”...
  • Page 19: Preliminary Operations

    8.1.2 Preliminary Operations Following are reported the parameters to set before performing the combustion analysis: 15/06/18 10:00 15/06/18 10:00 15/06/18 10:00 ANALYSIS [1/3] MENU CONFIGURATION 4.2 % ►Measurements ►Analysis 23 ppm Configuration Instrument 2.9 % Diagnostic Print ---% Service Info. Language Loss ---%...
  • Page 20: Performing The Combustion Analysis

    AVAILABLE IF THE VERSION OF THE INSTRUMENT Printer FORESEES IT . QR Code Pairing BT Print Print 15/06/18 10:00 WARNING Printing in progress ... SCAN THE QR CODE USING THE E INSTRUMENTS APP “E-Instruments QR Code App”, TO DOWNLOAD THE DATA.
  • Page 21: End Of The Analysis

    8.1.5 End of the Analysis - At the end of the combustion analysis, carefully remove the sample probe and remote air temperature probe, if used, from their relative ducts, taking care not to get burnt. - Switch off the instrument. Then, proceed to turn off the instrument.
  • Page 22 Cleaning hose Fig. b Maintaining the water trap / filter unit To remove the water trap, just rotate the cover and unhook the filter holder body; remove the internal cup and then replace the filter (see figure on the side). Clean all the filter parts using water only, dry the components and reassemble the filter.
  • Page 23: Menu

    9.0 MENU 9.1 Menu 15/06/18 10:00 FUNCTION MENU Returns to the previous screen. ►Measurements Configuration Diagnostic Selects the available parameters. Info service Enters in the selected parameter setting. SUB-MENU FUNCTION Through this menu, it is possible to execute the draft, ambient CO and pressure Measurements.
  • Page 24: Menu→Measurements

    Menu→Measurements 15/06/18 10:00 FUNCTION MEASUREMENTS Returns to the previous screen. ►Draft CO air Pressure Selects the available parameters. Enters in the selected parameter setting. SUB MENU FUNCTION The DRAFT menu gives access to the stack draft measurement. NOTE: The measurement may not be accurate due to condensation inside the gas probe.
  • Page 25: Menu→Measurements→Draft

    9.2.1 Menu→Measurements→Draft FUNCTION 15/06/18 10:00 DRAFT Activate the context keys shown on the Inlet display. Draft 0.00 h ►Zero sensor Returns to the previous screen. Starts the pressure sensor autozero. Save Print INTERACTIVE OPERATION FUNCTION The measure will be printed on the ticket of the current combustion analysis Save According to the version of the instrument and/or according with the related Print...
  • Page 26: Menu→Measurements→Co Air

    9.2.2 Menu→Measurements→CO air FUNCTION 15/06/18 10:00 CO AIR Activate the context keys shown on the display. 412 p CO Max 413 p Returns to the previous screen. Save Print INTERACTIVE OPERATION FUNCTION The measure will be printed on the ticket of the current combustion analysis Save According to the version of the instrument and/or according with the related Print...
  • Page 27: Menu→Measurements→Pressure

    9.2.3 Menu→Measurements→Pressure FUNCTION 15/06/18 10:00 PRESSURE Activate the context keys shown on the display. Press. 0.00 h ►Zero Sensor. Returns to the previous screen. Perform the Zero Sensor of the pres- sure sensor. Save Print INTERACTIVE OPERATION FUNCTION The measure will be printed on the ticket of the current combustion analysis Save According to the version of the instrument and/or according with the related Print...
  • Page 28: Menu→Configuration

    Menu→Configuration FUNCTION 15/06/18 10:00 CONFIGURATION Returns to the previous screen. ►Analysis Instrument Alarm Selects the available parameters. Print Language Reset Enters in the selected parameter. SUB MENU FUNCTION Through this menu the user, can set the different reference parameters of the instrument in order to perform the combustion analysis.
  • Page 29: Menu→Configuration→Analysis

    9.4.1 Menu→Configuration→Analysis FUNCTION 15/06/18 10:00 ANALYSIS Returns to the previous screen. ►Fuel Condensation reference Selects the available parameters. Measure units Autozero Air temp. Enters in the selected parameter. SUB MENU FUNCTION Lets the user select the type of fuel to be used during analysis. Fuel SEE CHAPTER 9.4.1.1.
  • Page 30: Menu→Configuration→Analysis→Fuel

    9.4.1.1 Menu→Configuration→Analysis→Fuel 15/06/18 10:00 15/06/18 10:00 FUEL [1/2] FUEL [2/2] ►Natural gas Coal Propane ►Bio-fuel 5% #2 Oil Bagasse #4 Oil L.P.G. #6 Oil Butane Diesel Wood/Pel. 8% FUNCTION Returns to the previous screen. Scrolls through the pages. Confirms the choice of fuel to be used during the analysis.
  • Page 31: Menu→Configuration→Analysis→Condensation

    9.4.1.2 Menu→Configuration→Analysis→Condensation 15/06/18 10:00 CONDENSATION Altitude above sea level ►Altitude Relative humidity of air R.H. air FUNCTION When pressed in modify mode cancels the selection made, otherwise re- turns to the previous screen. The arrows select each line displayed. In edit mode, it scrolls through the suggested values.
  • Page 32: Menu→Configuration→Analysis→O2 Reference

    9.4.1.3 Menu→Configuration→Analysis→O2 Reference 15/06/18 10:00 02 REFERENCE Percentage of Oxygen in CO measurement ►CO 0.0 % FUNCTION When pressed in modify mode cancels the selection made, otherwise returns to the previous screen. The arrows select each line displayed. In edit mode, it scrolls through the suggested values.
  • Page 33: Menu→Configuration→Analysis→Measure Units

    9.4.1.4 Menu→Configuration→Analysis→Measure units 15/06/18 10:00 MEASURE UNITS ►CO Measurement unit can be set as: ppm - mg/m - mg/kWh - g/GJ - g/m ng/J - g/kWh - % Pressure Measurement unit can be set as: ppm - mg/m - mg/kWh - g/GJ - g/m ng/J - g/kWh - % Measurement unit can be set as: hPa - Pa - mbar - mmH O - mmHg - inH...
  • Page 34: Menu→Configuration→Analysis→Autozero

    9.4.1.5 Menu→Configuration→Analysis→Autozero 15/06/18 10:00 AUTOZERO ►Autozero 30 s Duration of autozero, expressed in seconds. Purging Duration of the cleaning cycle, expressed in seconds. FUNCTION When pressed in modify mode cancels the selection made, otherwise re- turns to the previous screen. >...
  • Page 35 15/06/18 10:00 AUTOZERO ►Autozero Purging...
  • Page 36: Menu→Configuration→Analysis→Air Temp

    9.4.1.6 Menu→Configuration→Analysis→Air temp 15/06/18 10:00 AIR TEMP. T probe --- F ►T air 32.0 F T probe FUNCTION Enters edit mode of the selected element and then confirms the change. > When in modify mode, sets the desired value. When pressed in modify mode cancels the selection made, otherwise re- turns to the previous screen.
  • Page 37: Menu→Configuration→Analysis→Nox/No Ratio

    9.4.1.7 Menu→Configuration→Analysis→NOx/NO Ratio (if the version of the instrument foresees it) 15/06/18 10:00 NOx/NO RATIO ►NOx/NO 1.05 FUNCTION Enters edit mode of the selected element and then confirms the change. > When in modify mode, sets the desired value. When pressed in modify mode cancels the selection made, otherwise re- turns to the previous screen.
  • Page 38: Menu→Configuration→Instrument

    9.4.2 Menu→Configuration→Instrument 15/06/18 10:00 FUNCTION INSTRUMENT Returns to the previous screen. ►Display On site calib. Clock Selects the available parameters. Bluetooth Enters in the selected parameter. SUB MENU FUNCTION With the arrow keys it is possible to increase or decrease the brightness and the contrast of the display.
  • Page 39: Menu→Configuration→Instrument→Display

    9.4.2.1 Menu→Configuration→Instrument→Display 15/06/18 10:00 DISPLAY ►Brightness Contrast FUNCTION When pressed in modify mode cancels the selection made, otherwise re- turns to the previous screen. The arrows select each line displayed. In edit mode, it scrolls through the suggested values. Enters the modify mode for the selected parameter, then confirms the modification.
  • Page 40: Menu→Configuration→Instrument→On Site Calib

    9.4.2.2 Menu→Configuration→Instrument→On site calib. 15/06/18 10:00 ON SITE CALIB. FUNCTION Sets the password. Selects line; in modification sets the value or the desired mode. Once password is entered, gives access to the 'On site calibration' menu. Returns to the previous screen. When in modify mode cancels the modification just made.
  • Page 41 1. Start the instrument 15/06/18 10:00 15/06/18 10:00 AUTOZERO AUTOZERO AutoZero in Fresh T: 21.5°C Keep pressed E500 for a few S.n.:0 seconds Fw:1.00 AutoZero in Fresh Keep Repeat WARNING Make sure autozero is execute in clean air and terminates correctly.
  • Page 42 3. Once in the ‘On site calibration’ menu, is shown the list of the installed sensors for which the recalibration is available. In the recalibration screen all information related to the last performed calibration is shown, as well as the relevant values. Calibrate: saves new calibration 15/06/18...
  • Page 43 0,5 l/m...
  • Page 44 The calibration will be possible only when the status is set to ' ---- ' or 'inactive'. • 15/06/18 10:00 15/06/18 10:00 15/06/18 10:00 ON SITE CALIB. ON SITE CALIB. ON SITE CALIB. Calibrate Calibrate Calibrate ►Status ►Status active ►Status inactive El.time El.time...
  • Page 45 After the stabilization time, select 'Calibrate' and activate the function ' ' to store the new calibration. • Messages in the 'Status' line: 15/06/18 10:00 15/06/18 10:00 saving: the instrument is saving the performed ON SITE CALIB. ON SITE CALIB. calibration error: the sensor has NOT been recalibrated ►Calibrate...
  • Page 46: Menu→Configuration→Instrument→Clock

    9.4.2.3 Menu→Configuration→Instrument→Clock 15/06/18 10:00 CLOCK Time 10:00 Time, in the chosen format Date 15/06/18 Date, in the chosen format Date format: EU (Europe) or USA (America) Mode Time format: 24h or 12h Mode FUNCTION When pressed in modify mode cancels the selection made, otherwise returns to the previous screen.
  • Page 47: Menu→Configuration→Instrument→Bluetooth

    9.4.2.4 Menu→Configuration→Instrument→Bluetooth (if the instrument version foresees it) 15/06/18 10:00 BLUETOOTH Bluetooth status: On or Off Status Device ID Device MAC address 008025CE0E6E FUNCTION Activate the context keys shown on the display. Goes back to the previous screen. INTERACTIVE OPERATIONS DESCRIPTION Turns off Bluetooth ®...
  • Page 48: Menu→Configuration→Alarm

    9.4.3 Menu→Configuration→Alarm 9.4.3 Menu→Configuration→Alarm 11/12/17 10:00 ALARM ►Measure Observed parameter: CO - NO - O2 Mode Kind of set alarm: Max / Min / Off Limit 18.0 % Concentration threshold of the observed gas. FUNCTION Enters the modify mode for the selected parameter, then confirms the modification.
  • Page 49: Menu→Configuration→Print

    QR code generation (if the instrument version foresees it): ON: pushing the button related to the interactive function “Print” the instrument generates a QR code, which can be read with the “E Instruments Analyzer App” QR code and allows to download the data acquired of the performed combustion analysis and the additional measures.
  • Page 50: Menu→Configuration→Print→Pairing

    9.4.4.1 Menu→Configuration→Print→Pairing 1. When the Bluetooth printer is set, proceed with the following procedure: 15/06/18 10:00 15/06/18 10:00 15/06/18 10:00 PAIRING BT PAIRING BT PRINT Printer Turn on the Copy number printer and detection in Printer progress ... start searching Start QR code ►Pairing BT...
  • Page 51: Menu→Configuration→Language

    9.4.5 Menu→Configuration→Language 15/06/18 10:00 15/06/18 10:00 LANGUAGE [1/2] LANGUAGE [2/2] ►Italiano √ ►Polski English Pусский Español Français Hrvatski Deutsch FUNCTION Returns to the previous screen. Scrolls through the available languages. Sets the selected language. Example: 15/06/18 10:00 15/06/18 10:00 LANGUAGE [1/2] LANGUAGE [1/2] ►Italiano √...
  • Page 52: Menu→Configuration→Reset

    9.4.6 Menu→Configuration→Reset 15/06/18 10:00 15/06/18 10:00 RESET RESET Deleting and Clear memory data resetting data and restore factory setting? Push OK to continue FUNCTION Exits the current screen without resetting. Starts the factory data reset phase.
  • Page 53: Menu→Diagnostic

    Menu→Diagnostic FUNCTION 15/06/18 10:00 DIAGNOSTIC Returns to the previous screen. ►Sensors Gas probe Hardware Selects the available parameters. Enters in the selected parameter. SUB MENU DESCRIPTION Displays information on the state and calibration of the electrochemical sensors. Moreover, accesses to the ID data of the sensor: Code Serial number Sensors...
  • Page 54: Menu→Diagnostic→Sensors

    9.5.1 Menu→Diagnostic→Sensors Messages on the state and calibration of the electrochemical sensors: 15/06/18 10:00 SENSORS No problem detected absent The sensor was not detected ►S1: O2 err data Memory data error of the sensor S2: CO unknown It is necessary to update the FW of the device err pos The sensor has been installed in the wrong position S3: NO...
  • Page 55: Menu→Diagnostic→Gas Probe

    9.5.2 Menu→Diagnostic→Gas probe Connect flue sampling probe filter unit 15/06/18 10:00 assembly to the instrument; GAS PROBE Fully insert the black rubber cap on the gas probe tip, as shown in the following picture: Close the flue gas probe Provided black rubber cap Push OK to continue FUNCTION...
  • Page 56: Menu→Diagnostic→Hardware

    9.5.3 Menu→Diagnostic→Hardware FUNCTION 15/06/18 10:00 HARDWARE Returns to the previous screen. Memory Calibration ►Voltages Selects the available parameters. ADC channels Enters in the selected parameter. INTERACTIVE OPERATIONS DESCRIPTION Shows the values in mV. Shows the values in Bit. Note: the memory and calibration parameters are not accessible if their condition is OK. Visualization example: 11/12/17 10:00...
  • Page 57: Menu→Info Service

    Menu→Info Service 09/05/19 10:00 Info ser. [1/2] E instruments Tel. +33 (0)5 53 80 85 00 Fax. +33 (0)5 53 80 16 81 www.e-inst.com 09/05/19 10:00 Info ser. Model S.N. 9999 9.99 FW rev. FW P.N. HW rev. 1.00 Boot rev.
  • Page 58: Sensors

    Sensor characteristics diminish as the reagents are consumed and when these have been used up completely the sensor must be replaced. The sensors must be recalibrated on a regular basis to assure measuring accuracy: recalibration can only be performed by a qualified E INSTRUMENTS service centre. Table 10.4 illustrates the characteristics inherent to each sensor.
  • Page 59: Maintenance

    At least once a year send the instrument to a SERVICE CENTER for a complete overhaul and thorough internal cleaning. E INSTRUMENTS’ highly qualified staff is always at your disposal and will provide you with all the sales, technical, application and maintenance details required.
  • Page 60: Replacing The Gas Sensors

    11.3 Replacing the gas sensors The gas sensors of the instrument shall be periodically replaced with new or recalibrated sensors. The user can easily perform this replacement operation according to the following instructions: Before proceeding with the replacement of the sensor, it is necessary to insert in the instrument the ID code of the new sensor, after installing the appropriate PC software downloadable from the website www.e-inst.com.
  • Page 62 Insert the new sensor paying attention to match the sensor connectors with the relevant supports on the circuit board. Place back the tubes (See point 4). Close back the base of the instrument and screw back the four screws (See point 2). Turn on the instrument to check the new sensor works correctly through the menu "Sensor Troubleshooting".
  • Page 63: Replacing The Battery Pack

    11.4 Replacing the battery pack Follow these instructions to replace the battery pack: Extract the battery pack. Remove the base of the instrument Remove the battery pack connector, and replace the pack with a new one following the reverse procedure described above.
  • Page 64 Battery pack connector ONCE THE BATTERY IS REPLACED IT IS NECESSARY TO RESET DATE AND TIME THROUGH THE PARAMETER “CLOCK” (SEE CHAPTER 9.4.2.3).
  • Page 65: Troubleshooting

    That battery pack is old. Aging can cause the batteries to reduce their capacity. If the autonomy has become unacceptable change the internal battery The instrument battery autonomy is lower than what with an original part E INSTRUMENTS. stated in the “Technical features” chapter. d. Verify measured...
  • Page 66 Troubleshooting guide SYMPTOM PROBABLE CAUSES AND REMEDIES The following symbol "----" appears on the analysis The instrument is not able to calculate a numerical value screen. based on the flue gas analysis conducted. The "----" are replaced by numbers when the analyzer detects valid combustion data.
  • Page 67: Spare Parts And Servicing

    AAC SE58 CO Sensor with NOx filter 4000ppm, pre-calibrated and interchangeable 13.2 Service Centers To contact the Service center : E Instruments Zone Industrielle - BP16 24700 Montpon FRANCE Tel. +33(0)5 53 80 85 00 Fax. +33(0)5 53 80 16 81...
  • Page 68 Select the data separation mode: comma (,) or semicolon (;). This setting is useful if it’s needed to import the csv file in an electronic sheet such as Excel or Google Drive. Visualize the version of the APP and E Instruments contacts.
  • Page 69 Select one of the two modes of data sharing. Select the file format to share. Select the application to use for sending. Example of the exported csv file and imported in an Excel file: E500 Serial num. 1100 Date 22/12/2017 Time 12:00...
  • Page 70 ANNEX B Heading of the printed report with the PC Software “Easy2print” The ticket heading is easily settable by the user following the simple instructions below. 1. Connect the analyzer to the PC (with Windows XP operative system or later) through the USB cable. 2.
  • Page 71 ANNEX C Optional Measurements list: MEASURE DEFINITION Air excess. Expressed as a percentage according to the formula in the annex C, is the e (Exc. Air) ratio between the volume of air actually entering the combustion chamber and the one theoretically needed.
  • Page 72: Measurement Units Matching → Abbreviations

    MEASURE DEFINITION Total efficiency in relation to the Higher Heating Value: ηt (HHV) Total efficiency. It is the sum of sensible efficiency and condensation efficiency. It is re- ferred to HHV (Higher Heating Value) and can not exceed 100%. Total stack losses: Qt (HHV) It is the total heat percentage dissipated through the stack.
  • Page 73 ANNEX D Coefficients of the fuels and Formulas The following chart, lists the coefficients of the memorised fuels, used for calculating losses and efficiencies. Details of the coefficients of the fuels: Coefficients for calculating combustion efficiency CO2t M air V dry gas Fuel A1 USA (KJ/Kg)
  • Page 74 ANNEX E DECLARATION OF CONFORMITY The manufacturer: E Instruments by Sauermann with registered address in: KIMO S.A. - CS 5001 - 24700 MONTPON - FRANCE declares that the following products: E500 is in conformity with the essential requirements of directives 2014/30/UE and 2014/35/CE.