Configuration=>Diagnostic=>Gas Probe - E Instruments E4500 Operating & Maintenance Manual

Combustion analyzer
Hide thumbs Also See for E4500:
Table of Contents

Advertisement

9.7.2 Configuration→Diagnostic→Gas probe
07/08/14
10:00
Diagnostic
Gas probe
Close the flue gas probe
Press OK to start
OK
KEY
CONTEXT KEY
OK
Tightness test of the probe.
07/08/14
10:00
Diagnostic
Gas probe
Close the flue gas probe
Press OK to start
OK
Results:
Tightness: The system is OK
Error:
Make sure that the probe is connected to the input P-, check the seals of the pneumatic connections
and/or the seal of the condensation trap and check that the test cap is correctly inserted on the tip of
the probe. WARNING: a damaged probe tip may impair the test.
Connect the flue gas sampling probe and water trap assembly to the
instrument;
Fully insert the black rubber cap on the gas probe tip, as shown in
the following picture:
FUNCTION
Activate the context keys shown on the display.
Activates the context key located in the left side of the display.
Returns to the previous screen.
FUNCTION
Starts the test to check the tightness of the gas sampling probe.
Starts the test of the gas sampling probe.
Diagnostic
Gas probe
Close the flue gas probe
Press OK to start
Calibration
OK
Black rubber cap
07/08/14
10:00
Diagnostic
Gas probe
Close the flue gas probe
Press OK to start
Calibration
Probe test
50
07/08/14
10:00
Diagnostic
Gas probe
Close the flue gas probe
Press OK to start
Calibration
Probe test
Result: leak
K000000000EJ 023826A0 100714