Troubleshooting
Byte
Value Error description
E7
0x80
Parameters for measuring mode
switches are not supported
E8
0x01
No extension block present in the
sensor
E8
0x01
The sensor transmits its fault sig-
nal
E8
0x02
Sensor test failed
E8
0x02
Basic initialization required
E8
0x02
Inconsistency of the analog signal
E8
0x04
Potentiostat voltage cannot be
read
E8
0x04
Supply current is not OK
E8
0x08
Internal potentiostat voltage can-
not be read
E8
0x08
Clean mirror
E8
0x08
Signal outside the permissible
range
E8
0x10
Too many lookup values in the
sensor
E8
0x10
Analog pellistor voltage outside the
permissible range
E8
0x20
Sensor EEPROM not supported
E8
0x20
Reconfigure the gas
E8
0x40
SIOS cannot process sensor test
parameters
E8
0x40
Terminal voltage is not OK for the
PIR 7x00
E8
0x80
Sensor test failed, sensor can be
refilled
E8
0x80
ICOM and 4 to 20 mA interfaces of
the PIR 7x00 transmit different val-
ues.
E9
0x01
Incorrect sensor part number (sen-
sor lock activated)
E9
0x02
Compulsory sensor test during
startup failed
E9
0x04
Reference electrode test failed
E9
0x08
An error occurred when filling the
sensor data buffer
E9
0x10
Communication error or timeout
during sensor test
E9
0x10
P 8000 cannot set the measuring
range of the PIR 7x00
36
Remedy
Replace sensor
Replace sensor
Replace sensor
Replace SIOS assembly
Disconnect and reconnect sensor
Replace SIOS assembly
Disconnect and reconnect sensor
Replace sensor
Replace sensor
Replace sensor
Refill electrolyte
Deactivate sensor lock or use a sensor with the
correct part number
Check connection to SIOS and sensor
Replace sensor
Disconnect and reconnect sensor
Replace sensor
Disconnect and reconnect sensor
Replace sensor
Disconnect and reconnect sensor
|
Technical Manual
Digital Communication - Modbus RTU