3 ELECTRICAL INSTALLATION

3.1 Bonding/Grounding – Figs. 3.1 to 3.9
Caution. For safety reasons and optimum performance, the flowmeter, pipelines and medium
must be correctly bonded and grounded according to regulations.
Notes.
• The flow sensor must not be connected to a ground spike.
• For bonding connections use 4mm
• Sensors 80mm do not require grounding rings.
• Connect the transmitter ground connection to the flowmeter body ground – see Figs 3.6 and
3.7.
Key for Figs. 3.1 to 3.9
M e t a l P i p e s
P l a s t i c P i p e s
Insulation
2
(<10AWG) cable.
Fluid Contact Rings
Fig. 3.1 All Metal Pipe, Including Lined Metal Pipe.
Fluid Contact Rings
Fig. 3.2 Metal Pipe With Flange Adaptor
See Figs. 3.6 and 3.7