N O T E
Should the characteristic bend at 2 levels as illustrated below,
adjust P1 so the highest level o f the waveform on Channel 2 i s higher
than that of Channel 1.
Adjust P2 and L2 as same as 3 to 6.
7. Remove the oscilloscope probes.
8. Reset the switch SW1 t o the NOR position (UP).
9. Replace the CCU cover making sure that the a i r vent i s located on the
left.
10. Adjust BOOSTER according t o chart (See next page.).