Hamamatsu Photonics C2400 Instruction Manual page 29

Microscope video camera
Hide thumbs Also See for C2400:

Advertisement

(5) I n s e r t the H2503 option card into the connector labeled JB BOOSTER.
(6) Replace the CCU cover making sure that the air vent i s located on the
left, and tighten the 4 holding screws.
Booster Value Adjustment
The BOOSTER option comes preset with a gamma value of 0.5 however this
value i s continuously adjustable from 0.3 to 1.0.
MA TERIALS :
A2-Channel Oscilloscope (20MHz or greater) with two 10x probes.
SET UP CONDITIONS
Set Channel 1 and 2 of the oscilloscope as listed below:
Volts Per Division : 0.1V
Seconds Per Division : 20 microseconds