Download Print this page

I N S T A L L A T I O N - Hamamatsu Photonics C2400 Instruction Manual

Microscope video camera
Hide thumbs Also See for C2400:

Advertisement

6 . I N S T A L L A T I O N

6-1 UNP ACKING
Carefully unpack the equipment, and see that a l l system components described
in Section 3 of this manual are secure and undamaged.
6-2 INSTALLATION SITE
W h e n installing the equipment, attention should be paid to the following
points:
1. Do not use or locate the equipment near a strong electrostatic or
magnetic field.
2. Situate where temperature i s in the range of 0 - 4 0 t .
3. The site chosen should be relatively free of temperature fluctuations.
4. Humidity should be maintained at less than 90% (non-condensing).
5. Location site should be free of vibration.
6. The instrument should not be exposed to highly corrosive gas(es).
(Chlorine, fluorine, etc.)
7. Keep location clean.
8. Avoid direct exposure to sunlight.
9. Allow adequate space around the CCU for efficient a i r exchange -exhaust
fan at rear of CCU and side vents.
6-3 INSTALLATION OF CAMERA HEAD
Connection of the camera head to the optical imaging system is made via a
standard, 1" C-mount adaptor.
Depending on the geometry of the camera head installation i t may be neces-
sary to stabilize the camera via a camera stand, base plate, etc.

Advertisement