Booster Value Adjustment :
INTENSITY CONVERSION i s adjusted using 4 pots.(P1, L l , OFS, GAIN). T h e
meaning of the pots. and adjusting techniques described in the following
steps.
1. Turning pots.(L1,P1,L2,P2) fully CW.
2. Using the GAIN pot., adjust the slope of the waveform at TP2
(Channel 2 ) . GAIN determines the slope of the characteristic.
Turning GAIN CW results in an increased slope, CCW in a decreased
slope. When gamma i s 0.5, the slope i s approximately 4.
3. Turn P1 f u l l y CCW.
4. Using Ll pot., adjust the bending level of the waveform on TP2
(Channel 2 ) . L l determines the bending level of the characteristic.
Turning Ll CW will set a higher level, CCW a lower level.
Bending-
level low
Bending level high