5. Using the OFS pot., adjust the OFFSET to make the OV level o f
Channel 1 and Channel 2 identical. Turning the OFS pot. determines
the OV DC level of the characteristic, turning CW results in the level
decreasing, CCW increasing.
level high z
level low
6. Using P1, adjust the waveform of Channel 2 to be the same height as
Channel 1. P 1 determines the t i p bending rate of the characteristic;
turning P1 CW results in a decreased rate, CCW an increased rate.