Dräger Polytron 5100 Instructions For Use Manual page 23

Table of Contents

Advertisement

www.norrscope.com
Calibration Messages
Display Sequence
Over range
Under range
High calibration
HI GAS
gas concentration
Low calibration
LO GAS
gas concentration
Dräger Polytron 5100
Fault
Zero Cal: Exceeding the lower zero calibration limit
Span Cal: Exceeding the lower span calibration limit
Zero Cal: Exceeding the upper zero calibration limit
Span Cal: Exceeding the upper span calibration limit
Gas concentration exceeds sensor range during span
calibration
Gas concentration too low to perform proper span
calibration
Cause
Maintenance
Remedy
Zero Cal: Press the down key
Span Cal: Press the down key
Zero Cal: Press the up key
Span Cal: Press the up key
Use lower cal gas concen-
tration or press down key
Use higher cal gas concen-
tration or press up key
23