Calibrating The Conductivity Meter - Miele PLW 8615 Operating Instructions Manual

Laboratory glassware washer
Hide thumbs Also See for PLW 8615:
Table of Contents

Advertisement

Calibrating the
conductivity me-
ter
When particularly high demands are placed on the cleaning and final
rinse result, e.g., in chemical analysis, the conductivity meter can be
Adjust conductivity module
calibrated with the
system is calibrated from DI water to cold water in the conductivity
measuring range.
Conductivity meter calibration can only be carried out if there are
no load carriers in the wash cabinet.
The wash cabinet should be rinsed with DI water to prevent high
measured values due to lye splitting.
Adjust conductivity module
Press the
The DI water measurement is carried out.
Please wait...
The message
Measurement active
the display, e.g.,
At the end of the measurement, the
displayed together with the input screen for the manually measured
value.
The infeed side door is released for the manual measurement.
Open the door and perform the manual conductivity measurement.
Close the door.
Press the button next to the
open the on-screen keyboard.
Enter the measured value using the on-screen keyboard and con-
firm with OK.
Measured values with decimal places need to be mathematically
rounded.
Press the OK button.
The cold water measurement is carried out.
Perform the same work steps as for DI water if the message
water conductivity: Actual
The values entered for
trol module.
Machine functions
The
You can cancel the conductivity meter calibration using the  but-
ton in the footer. The message
successful. Please repeat.
 button.
and the current process step appear in
.
DI water conductivity: Actual
DI water conductivity: New
is displayed.
Conductivity ... New
reappear in the display.
Adjustment of conductivity module un-
will appear on the display.
System settings
 function. The measuring
 parameter to
are taken over by the con-
is
Cold
67

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Plw 8616

Table of Contents