Carrying Out A Program Test - Miele PLW 8615 Operating Instructions Manual

Laboratory glassware washer
Hide thumbs Also See for PLW 8615:
Table of Contents

Advertisement

System settings
Carrying out a
program test
68
Newly created or modified reprocessing programs can be checked
Program test
with the
Checked
 parameter if a  reprocessing program has run through the
Program test
without any faults (see programming manual, "Program
administration").
Program test
The
can be used to check the cleaning performance.
During the program test, the program sequence stops after each pro-
gram phase and the door on the infeed side can be opened so that a
sample can be taken.
When you open the door between 2 program phases, please be
aware of the following sources of danger:
– risk of burning and scalding after high-temperature program
phases
– risk of contact with corrosive substances after program phases
that dispense process chemicals
– risk of inhaling toxic vapors after program phases that use disin-
fectants
Steam that escapes from the machine after high-temperature
phases can trigger a false alarm in smoke detectors.
 Universal
Program block
Temperature
Duration
After the  reprocessing program has started, an additional  button is
displayed next to the program phase. The  button has two functions:
- Continue the  reprocessing program without opening the door.
- Skip the next program phase.
Sampling example
Program test
Press the
The display shows the available load carrier options.
Press the button for the required load carrier.
The available reprocessing programs are displayed.
. A program can only be enabled with the
Main wash 1
 0 : 2 5
1 9 4 ° F
 button.
user/
BED
US
18.06.2018 15:05

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Plw 8616

Table of Contents