Maintenance measures

72

Maintenance

The machine should be serviced every 2,000 hours of operation, or
at least once a year, by Miele Technical Service.
Maintenance covers the following:
- electrical safety according to national regulations
- door mechanism and door seal
- any screw connections and connectors in the wash cabinet
- water inlet and drainage
- internal and external dispensing systems
- spray arms
- filter combination
- sump including drain pump and drain valve
- all mobile units, modules, and inserts (charged separately)
Where applicable:
- Steam condenser
- Drying unit
- conductivity monitoring
- Connected printer
- network connection
External documentation software and computer networks are not
tested by Miele Service.
The following operational tests are carried out as part of the mainte-
nance:
- a program sequence as a test run with thermoelectrical measure-
ment
- leak test on seals
- safety testing of all relevant measuring systems, including fault dis-
plays
- safety features