TDK-Lambda Genesys Series Manual page 34

Programmable dc power supplies, modbus tcp
Hide thumbs Also See for Genesys Series:
Table of Contents

Advertisement

*OPC
Parameter
To activate, write 0x0001 as parameter.
*OPT?
Returns
Returns 2 for MODBUS TCP interface.
*RST
Parameter
To activate, write 0x0001 as parameter.
*TRG
Parameter
To activate, write 0x0001 as parameter.
*WAI
Parameter
To activate, write 0x0001 as parameter.
ABORt
Parameter
To activate, write 0x0001 as parameter.
INITiate[:IMMediate]
Parameter
To activate, write 0x0001 as parameter.
INSTrument:COUPle
Parameter
NONE – write 0x0000; ALL - write 0x0001 as parameter.
GLOBAL:*RST
Parameter
To activate, write 0x0001 as parameter.
OUTPut:ENA:POLarity
Parameter
REV - 0x0000; NORM – 0x0001 as return value.
OUTPut:MODE?
Return
OFF - 0x0001; CV – 0x0002; CC - 0x0003; CP – 0x0004 as return value.
OUTPut:PROTection:CLEar
Parameter
To activate, write 0x0001 as parameter.
OUTPut:PROTection:FOLDback[:MODE]
Parameter
OFF - 0x0000; CC – 0x0001; CV - 0x0002 as return value.
OUTPut:PROTection:FOLDback:DELay
Unit
100mSec. Multiply register value by 100mSec. For example: 1 equals 100mec;
5 equals 500mSec.
30