METER SETUP

Setup mode allows the selection of the Temperature unit,
Auto‑off, Beep confirmation tone, the type of pH buffer set,
the Resolution and Information. To enter Setup mode press
and hold
button until "STD" and
by "SETUP" and
• "TEMP" is displayed on the
secondary LCD with the current
temperature unit (e.g. "TEMP °C"),
for °C/°F selection, use the SET
button. After the temperature unit
has been selected, press
"A-OFF" selection.
• Use the SET button, to navigate
through the auto‑off choices:
8 minutes ("8", default value), 60
minutes ("60") or disabled ("---").
Press
to confirm and to enter the
"BEEP" selection.
• To switch the beep tone ON or OFF,
press the SET button; press
confirm and to enter the calibration
buffer selection "pH 7.01 BUFF".
• To change the buffer set, the meter
will show the current buffer set: "pH
7.01 BUFF" (for standard buffer
set: 3.00/7.01/10.01) or "pH
6.86 BUFF" (for NIST buffer set:
3.00/6.86/9.18). Change the set with the SET button.
Press
to confirm and to enter pH resolution selection
"RESOL".
• To change the pH measurement
resolution between "0.1" and "0.01"
use the SET button; then press
confirm and to enter electrode calibration
information "INFO" selection.
14