Meter Setup - Hanna Instruments HI99163 Instruction Manual

Portable waterproof meat ph meter
Hide thumbs Also See for HI99163:
Table of Contents

Advertisement

Setup mode allows the selection of the Temperature unit,
Auto‑off, Beep, the type of pH buffer set, the Resolution and
Information. To enter Setup mode press and hold
until "STD" and
tag. Release the button.
• "TEMP" is displayed on the second‑
ary LCD with the current temperature
unit (E.g. "TEMP °C"), for °C/°F se‑
lection, use the SET button. After the
temperature unit has been selected,
press
"A-OFF" selection.
• Use the SET button, to cycle through
the auto‑off choices: 8 minutes ("8",
default value), 60 minutes ("60") or
disabled ("---"). Press
and to enter the "BEEP" selection.
• To switch ON or OFF the beep tone,
press the SET button; press
confirm and to enter the calibration
buffer selection "pH 7.01 BUFF".
• To change the type of calibration buf‑
fer set, the meter will show the current
buffer set: "pH 7.01 BUFF" (for stan‑
dard buffer set: 4.01/7.01/ 10.01)
or "pH 6.86 BUFF" (for NIST buffer
set: 4.01/6.86/ 9.18). Change the set
with the SET button. Press
firm and to enter pH resolution selection "RESOL".
• To change the pH measurement reso‑
lution between "0.1" and "0.01" use
the SET button; then press
firm and to enter electrode calibra‑
tion information "INFO" selection .
• To switch ON or OFF the electrode
condition indicator on the LCD, press
the SET button; press
options; Change the set with the SET
button, then press
to return to normal mode.
14
tag is replaced by "SETUP" and
to confirm and to enter the
to exit setup
to confirm and
to confirm
to
to con‑
to con‑
button