HI99111 Portable Waterproof pH/Temperature Meter for Wine Analysis...
Page 2
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
Notify your nearest Hanna Instruments Customer Service Center if damage is observed. Each HI99111 is delivered in a rugged carrying case and is supplied with: • FC10483 preamplified pH/temperature probe with glass body and DIN connector with 1 m (3.3’) cable...
It features a two button operation system and is simple to use. The HI99111 has a waterproof and compact casing, large dual‑line display, and automatic pH calibration at one or two points.
MAIN FEATURES • Simultaneous pH and temperature measurements on a large dual‑line LCD display • Automatic pH calibration at one or two points within two memorized buffer sets (standard or NIST) • Selectable temperature unit (°C or °F) • Electrode condition indicator •...
DISPLAY DESCRIPTION 1 Automatic Temperature Compensation indicator 2 Stability indicator 3 Battery percentage 4 Electrode condition indicator 5 Low battery indicator 6 Measurement unit 7 Primary LCD 8 Temperature unit 9 Secondary LCD 10 Meter modes indicator 11 pH calibration buffer(s) used OPERATIONAL GUIDE Each meter is supplied with batteries.
Page 9
TURNING THE METER ON To turn the meter ON, press the button on the front of the meter. If it does not turn on, make sure that the batteries are properly installed in their place. The meter is provided with an active acoustic signal when a key is pressed.
Page 10
→ Press any button to resume active measurements. ENTERING CALIBRATION MODE Press and hold the button until ”POWER” and is replaced by “STD“ and tag. Release the button. ENTERING SETUP MODE Press and hold button until “STD“ and is replaced by “SETUP“ and tag.
Page 11
For better accuracy, frequent calibration of the pH sensor with the meter is recommended. In addition, the meter must be recalibrated: a) whenever the pH electrode is replaced. b) after testing aggressive chemicals. c) when high accuracy is required. d) at least once a month. e) after cleaning the sensor.
Page 12
Two-point calibration Proceed with steps 1 through 3 under single‑point calibration using 7.01 (pH 6.86 for NIST) pH buffer first. Then follow steps below: The ”pH 3.00 USE” message is then displayed. Place the probe in the second calibration buffer (pH 3.00 or 10.01, or, if using NIST, pH 3.00 or 9.18).
Page 13
ELECTRODE CONDITION The display is provided with a probe icon (unless the feature is disabled from setup) which indicates the electrode status after calibration. The “condition” remains active for 12 hours (unless the batteries are removed). The electrode condition is evaluated only if the current calibration has two points.
METER SETUP Setup mode allows the selection of the Temperature unit, Auto‑off, Beep confirmation tone, the type of pH buffer set, the Resolution and Information. To enter Setup mode press and hold button until “STD“ and tag is replaced by “SETUP“ and tag.
• To switch the electrode condition indicator on the LCD ON or OFF, press the SET button; press to exit setup options. Change the set with the SET button, then press to confirm and to return to normal mode. BATTERY REPLACEMENT When the remaining battery life is less than 10% the battery tag blinks on the display to warn the user.
Code and serial number on the other side Reference junction Quick connector with Temperature sensor Rubber protection, and pH bulb 1 meter cable ELECTRODE MAINTENANCE PREPARATION • Remove the protective cap. Do not be alarmed if any salt deposits are present. Rinse with water. •...
Sensor check procedure on page 13. CERTIFICATION All Hanna Instruments conform to the CE European Directives. Disposal of Electrical & Electronic Equipment. The product should not be treated as household waste. Instead hand it over to the appropriate collection point for the recycling of electrical and electronic equipment which will conserve natural resources.
Technical Service department and then send it with shipping costs prepaid. When shipping any meter, make sure it is properly packed for complete protection. Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.