Specifications - Hanna Instruments HI99111 Instruction Manual

Portable waterproof ph/temperature meter for wine analysis
Hide thumbs Also See for HI99111:
Table of Contents

Advertisement

SPECIFICATIONS

‑2.00 to 16.00 pH / ‑2.0 to 16.0 pH
Range*
±825 mV (pH‑mV)
‑5.0 to 105.0 °C/23.0 to 221.0 °F
0.01 pH / 0.1 pH
Resolution
1 mV
0.1 °C/0.1 °F
±0.02 pH / ±0.1 pH
Accuracy
±1 mV (pH‑mV)
±0.5 °C up to 60 °C; ±1.0 °C outside
@ 25°C / 77°F
±1.0 °F up to 140 °F; ±2.0 °F outside
Temperature
Automatic
compensation
‑5.0 to 105.0 °C/23.0 to 221.0 °F
Automatic, 1 or 2 point selectable buffer set
pH calibration
Standard: 3.00; 7.01; 10.01 or
NIST: 3.00; 6.86; 9.18
FC10483
Probe (included)
temperature probe with glass body, DIN
connector and 1 m (3.3') cable
1.5V AAA (3 pcs.)
Battery type/life
approx. 1400 hours of continuous use
user selectable: after 8 min, 60 min or
Auto‑off
disabled
0 to 50 °C (32 to 122 °F)
Environment
RH max. 100%
Meter dimensions 154 x 63 x 30 mm (6.1 x 2.5 x 1.2")
Weight
196 g (6.91 oz.)
(with batteries)
Case ingress
IP67
protection rating
* the
FC10483
is limited to be used from 0 to 12 pH and from 0 to 80 °C
temperature (32 to 176 °F).