Voltage dips,
0% UT for 0.5 cycle,
short interruptions and
At 0°, 45°, 90°, 135°, 180°,
voltage variations on power
225°, 270° and 315°
supply input lines
0% UT; 1 cycle
IEC 61000-4-11
and 70% UT ; 25/30 cycles
Single phase at 0°
0% UT; 300 cycle
Power frequency (50/60 Hz)
30 A/m
magnetic field IEC 61000-4-8
50 Hz or 60 Hz
0% UT for 0.5 cycle,
Mains power quality should be that of a typical
At 0°, 45°, 90°, 135°, 180°,
commercial or hospital environment. If the user of the
225°, 270° and 315°
EQUIPMENT or SYSTEM requires continued operation
0% UT; 1 cycle
during power mains interruptions, it is recommended that
and 70% UT ; 25/30 cycles
the EQUIPMENT or SYSTEM be powered from an uninter-
Single phase at 0°
ruptible power supply or a battery.
0% UT; 300 cycle
30 A/m
Power frequency magnetic fields should be at levels
50 Hz and 60 Hz
characteristic of a typical location in a typical commercial
or hospital environment.
24 – 25