WIRELESS BLOOD PRESSURE MONITOR
Read this manual before use. Keep it for further reference.
Disclaimer: Information in this guide may change without notice.
The manufacturer assumes no responsibilities for errors that may appear in this guide.

Advertisement

Table of Contents
loading

Summary of Contents for Withings BP-801

  • Page 1 WIRELESS BLOOD PRESSURE MONITOR Read this manual before use. Keep it for further reference. Disclaimer: Information in this guide may change without notice. The manufacturer assumes no responsibilities for errors that may appear in this guide.
  • Page 3: Package Content

    Consult your doctor during pregnancy, or if you suffer from arrhyth- mia and arteriosclerosis. Please read this section carefully before using the Withings Wireless blood pressure monitor. Intended use The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from 9 inches to 17 inches (22cm- 42cm).
  • Page 4: Replacing The Aaa Alkaline Cells

    Replacing the AAA alkaline cells – Do not leave the blood pressure monitor unattended with If the low battery symbol appears in the Withings application, infants or persons who cannot express their consent. replace all four AAA alkaline cells at the same time.
  • Page 5: After Use

    AFTER USE Cleaning Maintenance – Do not use an alcoholic-base or solvent agent to clean the device If you can’t fix the problems using the troubleshooting instructions , – Clean the device with a soft and dry cloth request service from your dealer. Manufacturer will make available –...
  • Page 6: Error Messages

    ERROR MESSAGES Error message Causes Countermeasures Measurement could not be performed. Please try Hardware auto- ‐check error. 1- Measurement could not be performed. again. If the problem occurs again, please contact Please try again. If the problem occurs again, customer service. please contact customer service.
  • Page 7 Error message Causes Countermeasures Check that the Blood Pressure Monitor is - User movement, or not relaxed, or talking … 1- Check that the Blood Pressure Monitor is correctly positioned on your arm and that mea- - Cuff is not correctly fastened correctly positioned on your arm and that mea- surement is performed in good conditions.
  • Page 8: Troubleshooting

    TROUBLESHOOTING Problem Remedy Though the batteries are installed, there is nothing happening 1. Check that your iOS or Android device is unlocked when connected to iOS or Android device 2. Check and correct the AAA alkaline cells polarities 3. Remove the AAA alkaline cells and wait for one minute. Then install the AAA alkaline cells back.
  • Page 9 Problem Remedy Under normal measuring circumstance, the reading at home 1. The variation is due to the different environments is different from that of the clinics or each measurement has 2. The blood pressure is changing according to the physiological various reading or psychological status of the human body 3.
  • Page 10: Specifications

    SPECIFICATIONS Technical data Product description: Digital automatic Blood Pressure Monitor Model: BP-801 (Withings) Measurement method: Cuff oscillometric method Cuff inflation: Automatic inflation with air pump at 15 mmHg/s Pressure sensor: Gauge sensor Measurement range(pressure): 0 to 285 mmHg Measurement range (pulse):...
  • Page 11 Power source: AAA alkaline cells (x4) Weight: Approx. 600g without cells Accessories: AAA Alkaline cells x4, Micro-USB to USB adapter, Instruction manual Product life : 5 years Wireless transmission V4.0 dual mode (V2.1 + EDR) Note: Specifications are subject to change without prior notice or any obligation on the parts of the manufacturer...
  • Page 12: Reference To Standards

    REFERENCE TO STANDARDS This device complies with the following normative documents : automated non-invasive sphygmomanometers COUNCIL DIRECTIVE 93/42/EEC of 14 June 1993 concerning – EN300328 V1.8.1 Electromagnetic compatibility and Radio spectrum medical devices as amended by Directive 2007/47/EC Matters (ERM); Wideband transmission systems; Data transmission equipment operating in the 2,4 GHz ISM band and using wide band –...
  • Page 13 YA HORNG ELECTRONIC CO., LTD. ATTEN ELECTRONIC (DONGGUAN) CO., LTD FCC ID: XNAWPM02 188 industrial district, Ping Shan Administrative District, Tang Shia Town, Dongguan, Guangdong, China IP22 Ingress of water or particulate matter 70°C (158°F) -25°C (-13°F) WEEE Read this manual before use Kahl Handelsvertretung Isarstr.
  • Page 14: Fcc Statement

    FCC STATEMENT Federal Communications Commission (FCC) Statement 15.21 This device complies with Part 15 of the FCC Rules. You are cautioned that changes or modifications not expressly Operation is subject to the following two conditions: approved by the part responsible for compliance could void the user’s 1) this device may not cause harmful interference and authority to operate the equipment.
  • Page 15 - guidance power supply network that supplies buildings used for domestic purposes CE emissions The BP-801 Wireless Blood Pressure Group 1 – Portable and mobile RF communications equipment should be CISPR11...
  • Page 16 Declaration – electromagnetic emissions and immunity for equipment and systems that are not life-supporting and are specified for use only in a shielded location The Wireless Blood Pressure Monitor declaration electromagnetic immunity The Wireless Blood Pressure Monitor system is intended for use in the electromagnetic environment specified below. The customer or the user of the Wireless Blood Pressure Monitor system should assure that it is used in such an environment.
  • Page 17 Declaration – electromagnetic immunity The Wireless Blood Pressure Monitor system is intended for use in the electromagnetic environment specified below. The customer or the user of the Wireless Blood Pressure Monitor system should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level...
  • Page 18 DISPOSAL Actuation of European directives 2002/95/EC, 2002/96/EC and 2003/108/EC, for reduction in use of dangerous substances in the electric and electronic device and for garbage disposal. The symbol applied on the device or its packaging means that at the end of its useful life the product must not be disposed of with domestic waste.